RS232 communication Protocol

3.1 Transpor t protocols : 9600, 8, N, 1

    (1) single transmit data: Data receiver(PC) transmit “R”(52H)
          to DRO electronic box which well transmit data single. DRO
          electronic box receive each 52H, then transmit a set of data
          (current showing value).
    (2) Continuous transmit data: Data receiver(PC) transmit “U”(55H)
          to DRO electronic box which will transmit data continuous.
          It will transmit when ever showing value changed.
     (3)Stop to transmit data: DRO electronic box received 53H,
          then stop to transmit data. When enter in continuous 
          transmit data state, it must be transmit 53H to stop it, 
          then can receive other orders.
  

3.2 ASCII code number for mat:
 

 [Blank] X [sign-bit][Value][Value][Value][Value][Value][Decimal Point]

  [Value][Value][Value][Blank] Y [sign-bit][Value][Value][Value][Value]

  [Value][Decimal Point][Value][Value][Value][Blank] Z [sign-bit][Value]

  [Value][Value][Value][Value][Decimal Point][Value][Value][Value]

  [LF 0AH]

     

 Decimal point position unfixed, along with set-up resolution changed, 
 for example: 
  X+1234.567 (resolution = 1,2,5,10)  (Decimal point behind three figures)
  X+123.4567(resolution = 0.5, 0.1)  (Decimal point behind four figures)
  X+123.4567 (I/M =

 

I

 

INCH   I/M = M METRIC) 

                        (Inch more than one decimal point compare Metric)

3.3 Transmission line 1.5m wire, the following is 

       the connection method:

     PC connector                            DRO  connector

                  1                                                      1

     RX        2                                                RX  2

     TX        3                                                 TX 3

                  4                                                      4

     Gnd      5                                               Gnd 5

        

7

Summary of Contents for DS10-1V

Page 1: ...DS10 1V MultiFunction Version 2 0 ...

Page 2: ...not try to repair it please contact Delos or distributors Digital readout is a precise measuring device used with an optical Linear Scale When it is in use if the connection between the Linear Scale and the digital readout is broken or damaged externally incorrect measuring values may be resulted Therefore the user should be careful Do not try to repair or modify the digital readout otherwise fail...

Page 3: ...oggle Between R D display Mode 3 2 Parameters settings 2 1 PA 00 Encoder type setting 4 2 2 PA 01 Resolution setting 4 2 3 PA 02 Direction setting 4 2 4 PA 03 UART mode setting 4 2 5 PA 04 Accuracy of imperial unit 5 2 6 PA 05 Limit Enable disable 5 2 7 PA 06 Soft lower Limit value 5 2 8 PA 07 Soft Upper Limit value 6 2 9 PA 08 Restore Factory Settings 6 3 RS232 communication protocol 7 4 Relay ou...

Page 4: ...ee the processing operation and parameter programming 1 2 General Operations 1 Zeroing Press CLS Zeroing is to set the current point in normal display 2 Toggle Unit between inches and mm Press mm inch Display unit can be toggled between mm and inch a mm inch LED will be appear of the right at the same time Light indicator is the inches unit Otherwise is mm 3 Absolute lncremental Press ABS INC Disp...

Page 5: ...mpensation rectifying coefficient Measurement I Standard value x 1000 000 Coefficient Standard value Example for X axis Measurement 200 020mm Standard value 200 000mm Rectifying coefficient 200 020 200 1000 200 0 01 mm m 6 1 2 Function Set the center of work piece as datum by halving the displayed value Press SET then Press 1 2 in turn to value the axis display value The position is the work piece...

Page 6: ...10 0 for Linear Encoder and input value for Rotary Encoder Operations Windows displays PA 01 Press Enter key to enter For linear encoder Press CLS key to select the resolution For Rotary encoder Press SET key then press numeric number to input the parameter for rotary encoder when the input numeric number is positive the unit is degree when the input numeric number is negative the unit is degree m...

Page 7: ... Enter to confirm your choice and Windows displays PA 04 Press the MM INCH key to the last step Press the R D key to the next step 2 6 PA 05 Enable soft Limit function Parameters ON for enable or OFF for disable Operations Windows displays PA 05 Press Enter key to enter Press CLS key to toggle ON or OFF Press the Enter to confirm your choice and Windows displays PA 05 Press the MM INCH key to the ...

Page 8: ...rm your choice and Windows displays PA 07 Press the MM INCH key to the last step Press the R D key to the next step 2 9 PA 08 Restore factory default PINCoder 2019 Parameters Operations Windows displays PA 08 Press Enter key to enter Windows display PinCode input 2019 then press Enter Windows display CLS and then Windows displays PA 08 Press the MM INCH key to the last step Press the R D key to th...

Page 9: ... be transmit 53H to stop it then can receive other orders 3 2 ASCII code number format Blank X sign bit Value Value Value Value Value Decimal Point Value Value Value Blank Y sign bit Value Value Value Value Value Decimal Point Value Value Value Blank Z sign bit Value Value Value Value Value Decimal Point Value Value Value LF 0AH Decimal point position unfixed along with set up resolution changed f...

Page 10: ...4 Relay output 8 DSUB 9 1 2 3 4 5 6 7 8 9 pin 1 2 3 4 5 6 7 8 9 ON COM OFF EXP Color Red Yellow Black State Relay output DB9 ...

Page 11: ...D GND A A 5V 5V B B R R A B R Resolutions for Linear Scale Resolutions for Rotary Encoder Power Supply Axis Display Signal input per axis Maximum input frequency Operating Temperature Storage Temperature Relative Humidity Vibration Resistance Protection Class EN60529 Weight 10 0 1um 0 001 1 100VAC 230VAC 10 15 VA 7 Segment LED H 17mm A B quadrature signals 500KHz 0 50 20 70 95 not condensed 25 m s...

Page 12: ......

Reviews: