BD-2100 FEATURES
h
Designed to check moisture levels in wood, concrete, EIFS,
drywall, and other materials
h
Resistance technology recognized around the world as the
most accurate method for measuring moisture
h
Proven microcontroller circuit
h
6%-40% MC range on wood
h
0.2%-50% MC range on gypsum
h
0 to 100 reference scale for other non-wood materials
h
Easy one-hand operation
h
Includes (1) 9-Volt Battery
h
Includes sturdy carrying case
h
One-year warranty
Digital Readout
Read Key
Calibration
Check, Stats
Key
$
Set-Point
Key
#
Scale
Key
*
Color-coded
LED Indicators
&
Connector for
External
Electrode
@
#2497/A-100 Contact Pins
with 5/16” penetration
^
!
D E L M H O R S T B D - 2 1 0 0
2
The operating instructions for the model BD-2100 are divided
into two sections. Section A is a step-by-step guide to the
different features and basic use of your moisture meter.
Section B contains information on how the meter is used for
testing different materials.
Please read the instructions carefully. If you require additional
information, please contact Delmhorst for assistance.
SECTION A
BEFORE YOU BEGIN
KEY FUNCTIONS
!
Read Key
- This key reads the Percent Moisture Content
or relative moisture value in the material under test.
@
Calibration Check Key
- This key, (when pressed with the
read key) checks the meter calibration. It also displays
the number of readings in memory (up to 100), the
average, and the highest stored reading. It also clears
the memory.
#
Scale Key
- This key sets the display scale to #1 (Wood),
#2 (0-100 Reference scale), and #3 (Gypsum). It also
acts as a scroll key, depending on the function.
$
Set-Point Key
- This key programs the set-point value. A
buzzer will alert you if the meter reads higher than the
selected value. It also acts as a scroll key, depending on
the function.
CHECK CALIBRATION
h
Press the read key
!
and calibration check key
@
simultaneously. Meter is in calibration if it displays 12%
(+/- 0.2). Make sure the pins are not in contact with anything
when checking the calibration.
If you check the calibration and the meter does not display 12%,
it is likely an indication of a low battery. If this occurs, replace
the battery with a new one immediately. Continued use with a
low battery may cause the meter to go out of calibration. If you
have a fresh battery and the instrument still does not indicate a
correct calibration, return it to DELMHORST for service. See
Service for Your Meter
section.
Note:
When the battery is removed and then reconnected, the
meter displays its software version for one second and then
turns itself off. After replacing the battery, you must reset
the meter as described in
Resetting the Mete
r section.