Limited Warranty, Sole Remedy and Limited Liability

DeFelsko's sole warranty, remedy, and liability are the express limited
warranty, remedy, and limited liability that are set forth on its website:

www.defelsko.com/terms

Warning: The PosiTector PC Powder Checker contains a
very powerful magnetic mount for attaching the protective

rubber holster to nearby steel structures. As with all magnets,
avoid contact with electronic instruments or devices sensitive to
magnetic fields such as pace makers and coating thickness
gages. Care should also be given to the storing and shipping of
any strong magnet to prevent contact with sensitive devices.

Summary of Contents for PosiTector PC

Page 1: ......

Page 2: ...hout this manual the symbol indicates more information about the particular topic or feature is available on our website Go to www defelsko com manuals Power up Power down Step 1 Power up the probe by...

Page 3: ...m height 4 Release the button when you hear a double BEEP or see a thickness measurement on either display Both units will power down automatically after approximately 5 minutes of inactivity Measurem...

Page 4: ...ration Setting Unit of Measurement Maximum Amplitude Dot Probe Display button not pressed Amplitude Indicator Results Indicator Distance Target Bar Probe Batteries Status pg 15 Bluetooth Function Indi...

Page 5: ...ltrasonic echo A high position indicates good probe orientation relative to perpendicular The bar will generally reach higher positions when measuring thin powder and when the probe is at a 90 degree...

Page 6: ...Statistics Zero Cal Settings Setup Exit Appears when probe is ON Current selection is displayed with darkened background To navigate press to scroll DOWN to scroll UP and to SELECT To exit any menu pr...

Page 7: ...as the PosiTector 6000 Compare results A successful verification requires that the average of a series of measurements by both instruments be within their combined tolerances If not adjust the PosiTec...

Page 8: ...Adjustment is the act of aligning the gage s predictions to match those of a known cured sample Begin by performing a Verification of Accuracy pg 6 to determine if there is a need for an adjustment If...

Page 9: ...tolerances the PosiTector PC can be adjusted using one of three methods 1 1 point Adjustment The first and most common choice 2 2 point Adjustment If a 1 Pt Adjust does not improve accuracy 3 Powder...

Page 10: ...ess with a magnetic or eddy current coating thickness gage Record the results 10 Select 2 Pt Adjust then select the appropriate batch names when asked 11 Adjust the displayed batch averages lower or h...

Page 11: ...y default Cal 1 setting after a Reset pg 13 or if the user deletes all calibration settings below Creates a new calibration setting using the next available number Maximum of 9 By default these new Ca...

Page 12: ...d each measurement will now be simultaneously shown on the display and stored into this new batch On screen statistics are immediately updated with each measurement New batch names are date stamped at...

Page 13: ...ss both and buttons simultaneously to cancel printing Downloading Measurements A computer can view and download measurements stored in memory using the included USB cable Measurements are not erased f...

Page 14: ...t symbol appears It performs the same function as a menu Reset with the addition of Units microns Language English and Flip Display Normal A probe Zero pg 6 must be performed after any Reset This opti...

Page 15: ...e USB Drive A formatted HTML report is viewed by selecting the index html file found in the root directory Optionally readings0 txt files located in each batch folder i e B1 provide access to measurem...

Page 16: ...ed from the screen by no more than 0 2 MPa 20 30 psi of compressed air Do not direct air directly into the probe Replacement screens are available Changing the Batteries The battery symbol displays a...

Page 17: ...o com support Technical Data Conforms to ASTM D7378 Base unit dimensions 146 x 64 x 31 mm 5 75 x 2 5 x 1 2 Probe dimensions 222 x 50mm diameter 8 75 x 2 diameter Alkaline Batteries Base unit 3 AAA bat...