Introduction

The PosiTector

®

PC Powder Checker

®

is a hand-held, non-contact

Coating Thickness Gage that uses an airborne high frequency
ultrasound to analyze coating powder applied to rigid substrates to
calculate and display a predicted cured thickness.

The instrument consists of a base unit and a probe that communicate
with each other via Bluetooth wireless technology. They must stay
within 10 meters (30 feet) of each other while measuring. A rubber
holster with belt loop allows the base unit to stay with the operator while
measuring.

Note: Do not use compressed air on the probe. Tips for cleaning the
screen guard can be found on pg 15.

Note: Throughout this manual, the      symbol indicates more
information about the particular topic or feature is available on our
website. Go to: www.defelsko.com/manuals 

Power-up / Power-down

Step 1: Power-up the probe by pressing the      button.
Step 2: Power-up the base unit by pressing any one of the 3 buttons. A
beep and a solid probe icon on the base unit indicates both units are
ready to measure.

To preserve battery life, both units power-down after approximately 5
minutes of inactivity. All settings are retained.

When a USB cable is connected to the base unit, power is supplied
through the USB cable. The batteries are not used and the base unit
will not automatically power down. The probe will still automatically
power down.

The base unit can be powered down at any time by holding down both
(-) and (+) buttons simultaneously for 5 seconds. The probe will also
power down.

1

Summary of Contents for PosiTector PC

Page 1: ......

Page 2: ...hout this manual the symbol indicates more information about the particular topic or feature is available on our website Go to www defelsko com manuals Power up Power down Step 1 Power up the probe by...

Page 3: ...m height 4 Release the button when you hear a double BEEP or see a thickness measurement on either display Both units will power down automatically after approximately 5 minutes of inactivity Measurem...

Page 4: ...ration Setting Unit of Measurement Maximum Amplitude Dot Probe Display button not pressed Amplitude Indicator Results Indicator Distance Target Bar Probe Batteries Status pg 15 Bluetooth Function Indi...

Page 5: ...ltrasonic echo A high position indicates good probe orientation relative to perpendicular The bar will generally reach higher positions when measuring thin powder and when the probe is at a 90 degree...

Page 6: ...Statistics Zero Cal Settings Setup Exit Appears when probe is ON Current selection is displayed with darkened background To navigate press to scroll DOWN to scroll UP and to SELECT To exit any menu pr...

Page 7: ...as the PosiTector 6000 Compare results A successful verification requires that the average of a series of measurements by both instruments be within their combined tolerances If not adjust the PosiTec...

Page 8: ...Adjustment is the act of aligning the gage s predictions to match those of a known cured sample Begin by performing a Verification of Accuracy pg 6 to determine if there is a need for an adjustment If...

Page 9: ...tolerances the PosiTector PC can be adjusted using one of three methods 1 1 point Adjustment The first and most common choice 2 2 point Adjustment If a 1 Pt Adjust does not improve accuracy 3 Powder...

Page 10: ...ess with a magnetic or eddy current coating thickness gage Record the results 10 Select 2 Pt Adjust then select the appropriate batch names when asked 11 Adjust the displayed batch averages lower or h...

Page 11: ...y default Cal 1 setting after a Reset pg 13 or if the user deletes all calibration settings below Creates a new calibration setting using the next available number Maximum of 9 By default these new Ca...

Page 12: ...d each measurement will now be simultaneously shown on the display and stored into this new batch On screen statistics are immediately updated with each measurement New batch names are date stamped at...

Page 13: ...ss both and buttons simultaneously to cancel printing Downloading Measurements A computer can view and download measurements stored in memory using the included USB cable Measurements are not erased f...

Page 14: ...t symbol appears It performs the same function as a menu Reset with the addition of Units microns Language English and Flip Display Normal A probe Zero pg 6 must be performed after any Reset This opti...

Page 15: ...e USB Drive A formatted HTML report is viewed by selecting the index html file found in the root directory Optionally readings0 txt files located in each batch folder i e B1 provide access to measurem...

Page 16: ...ed from the screen by no more than 0 2 MPa 20 30 psi of compressed air Do not direct air directly into the probe Replacement screens are available Changing the Batteries The battery symbol displays a...

Page 17: ...o com support Technical Data Conforms to ASTM D7378 Base unit dimensions 146 x 64 x 31 mm 5 75 x 2 5 x 1 2 Probe dimensions 222 x 50mm diameter 8 75 x 2 diameter Alkaline Batteries Base unit 3 AAA bat...

Page 18: ...Warning The PosiTector PC Powder Checker contains a very powerful magnetic mount for attaching the protective rubber holster to nearby steel structures As with all magnets avoid contact with electroni...

Page 19: ......

Page 20: ......

Reviews: