5
The absence of an arrow does not necessarily signify
stability. It may mean that the particular reading is undergoing
little to no change. Therefore, slow changes could occur without
being indicated by a trend arrow.
To exit Live Data view, select
View Log Data
from the menu.
NOTE:
Smart Trend
TM
Indicators
- identify rising, falling or stable
readings. The probe compares each measurement with recent
values to provide the current direction of change.
Logging Mode
In
Logging Mode
, the
PosiTector DPM L
probe will
automatically record datasets into memory at user selected time
intervals. To enable, select
Start Logging
from the
PosiTector
App
or
PosiTector Advanced
gage menu, then enter the desired
logging measurement interval (between 1 minute and 8 hours)
and select
OK
. All previously logged datasets are deleted from
the probe’s memory.
The
PosiTector DPM L
immediately records the first dataset and
records subsequent datasets at the user-specified interval. The
PosiTector DPM L
does not need to remain connected to the
PosiTector App
or
PosiTector Advanced
gage body to continue
logging datasets. A maximum of 10,000 datasets (Ta, Ts, Td, Ts-
Td, RH, Tw) can be stored in
PosiTector DPM L
memory. All datasets are date and time stamped.
The
Logging Mode
icon will appear on the
display alongside the set interval. Measurements
on the display are only updated when a
dataset is stored. To view current
environmental
conditions, select
View Live Data
from
the menu. In View
Live Data mode, the
PosiTector DPM L
probe will continue
recording. To return
to Log view, select
View Log Data
.