background image

After the last measurement, the Gage will calculate a Zero which represents the
average of all the 

Zn Zero

readings taken.

N 1 Pt Adjustment 

For rough zinc surfaces (i.e. zinc spray metallizing), it may be desirable to adjust the
Gage to a known thickness, such as a shim placed over the zinc, rather than adjusting
it to zero. This ensures the Gage measures the thickness of paint over the metallizing
“peaks”.
1) Select 

N 1 Pt Adjust

from the 

Cal Settings

menu.

2) Press 

(+)

to select the number of readings to be used to obtain an average, typically

3 to 10 readings. The greater the variation between readings, the more readings should
be taken to obtain an average.
3) Repeatedly measure the known thickness reference on the unpainted galvanized
part. The Gage will wait two seconds between readings to allow the user to correctly
position the probe on the surface. After the last measurement, the Gage will calculate
and display the reading which represents the average of all the measurements taken.
If the expected reading is not obtained (within tolerance), lift the probe from the surface
and adjust the reading down 

(-)

or up 

(+)

to the expected thickness and press        .

NOTE:

The reference thickness should be similar to the expected paint thickness.

© DeFelsko Corporation USA 2016    All Rights Reserved

For more detailed instructions, see the included PosiTector 

6000 Quick Guide.

PosiTector 

6000 FNDS

probes are compatible with all PosiTector Advanced

models (serial numbers 730,000+) and Standard models with color display.

Probe Specifications

Mils

Microns

Minimum Zinc Thickness

0.5 mil 

13 µm

Paint Layer Accuracy

+/- (0.05 mil + 1%) 

0-2 mils

+/- (0.1 mil + 1%) 

> 2 mils

+/- (1 µm + 1%) 

0-50 µm

+/- (2 µm + 1%) 

> 50 µm

Zinc Layer Accuracy

+/- (0.1 mil + 1%) 

0-2 mils

+/- (0.2 mil + 1%) 

> 2 mils

+/- (2 µm + 1%) 

0-50 µm

+/- (4  µm + 1%) 

> 50 µm

For full specifications refer to defelsko.com/p6000/p6000_probe_details.htm

Summary of Contents for PosiTector 6000 FNDS

Page 1: ...ither protection system used independently Duplex Mode In Duplex mode the Gage utilizes both magnetic ferrous and eddy current non ferrous principles simultaneously to calculate and display the individual paint and zinc layer thicknesses The magnetic principle is used to measure the combined paint zinc thickness over the ferrous substrate and the eddy current principle is used to measure the paint...

Page 2: ...AT LEAST 2 inches 5 cm between readings or leave the probe on the surface in the same location for continuous measurements Verify accuracy on reference standards at the beginning and the end of each shift or if the gage is dropped or suspected of giving erroneous readings When enabled default the Gage will display two measurement values as shown above To disable uncheck Duplex within the Setup Men...

Page 3: ...nt 1 Select F Zero from the Cal Settings menu 2 Press to select the number of readings to be used to obtain an average typically 3 to 10 readings The greater the variation between readings the more zero readings should be taken to obtain an average 3 Repeatedly measure the uncoated steel part The Gage will wait two seconds after placing the probe on the surface to allow the user to correctly posit...

Page 4: ...o correctly position the probe on the surface After the last measurement the Gage will calculate and display the reading which represents the average of all the measurements taken If the expected reading is not obtained within tolerance lift the probe from the surface and adjust the reading down or up to the expected thickness and press NOTE The reference thickness should be similar to the expecte...

Reviews: