background image

The first time a part is measured, and any time the process changes, check zero on
both the bare steel substrate and on the uncoated zinc surface.  Adjust if necessary
(see 

Calibration Adjustment

).

Calibration Adjustment

Adjustment, or calibration adjustment is the act of aligning the Gage’s thickness
readings to match that of a known sample in order to improve the effectiveness of the
Gage on a specific surface or in a specific portion of its measurement range.  

PosiTector 

6000 FNDS

probes are factory calibrated and perform an automatic self-

check each time a measurement is taken. For many hot-dip galvanized applications,
no further adjustment is necessary from the factory calibration. The     symbol
disappears whenever a calibration adjustment is made. 

F Zero Adjustment

Measure the uncoated steel substrate (if available). If the average of several “N”
readings is not within tolerance of “0”, perform an 

F Zero

adjustment.

1) Select 

F Zero

from the 

Cal Settings

menu.

2) Press 

(+)

to select the number of readings to be used to obtain an average, typically

3 to 10 readings. The greater the variation between readings, the more zero readings
should be taken to obtain an average.
3) Repeatedly measure the uncoated steel part. The Gage will wait two seconds after
placing the probe on the surface to allow the user to correctly position the probe. After
the last measurement, the Gage will calculate a Zero which represents the average of
all the 

F Zero

readings taken.

Zn Zero Adjustment

Measure the unpainted zinc surface (if available). If the average of several “N” readings
is not within tolerance of “0”, perform an 

Zn Zero

adjustment.

1) Select 

Zn Zero

from the 

Cal Settings

menu.

2) Press 

(+)

to select the number of readings to be used to obtain an average, typically

3 to 10 readings. The greater the variation between readings, the more readings should
be taken to obtain an average.
3) Repeatedly measure the unpainted galvanized part. The Gage will wait two seconds
after placing the probe on the surface to allow the user to correctly position the probe. 

Summary of Contents for PosiTector 6000 FNDS

Page 1: ...ither protection system used independently Duplex Mode In Duplex mode the Gage utilizes both magnetic ferrous and eddy current non ferrous principles simultaneously to calculate and display the individual paint and zinc layer thicknesses The magnetic principle is used to measure the combined paint zinc thickness over the ferrous substrate and the eddy current principle is used to measure the paint...

Page 2: ...AT LEAST 2 inches 5 cm between readings or leave the probe on the surface in the same location for continuous measurements Verify accuracy on reference standards at the beginning and the end of each shift or if the gage is dropped or suspected of giving erroneous readings When enabled default the Gage will display two measurement values as shown above To disable uncheck Duplex within the Setup Men...

Page 3: ...nt 1 Select F Zero from the Cal Settings menu 2 Press to select the number of readings to be used to obtain an average typically 3 to 10 readings The greater the variation between readings the more zero readings should be taken to obtain an average 3 Repeatedly measure the uncoated steel part The Gage will wait two seconds after placing the probe on the surface to allow the user to correctly posit...

Page 4: ...o correctly position the probe on the surface After the last measurement the Gage will calculate and display the reading which represents the average of all the measurements taken If the expected reading is not obtained within tolerance lift the probe from the surface and adjust the reading down or up to the expected thickness and press NOTE The reference thickness should be similar to the expecte...

Reviews: