How to Measure

The 

PosiTector 

6000 

powers up when the center navigation button         is pressed.

To preserve battery life, the instrument powers down after approximately 5 minutes 
of no activity. All settings are retained.

1. Remove the protective rubber cap from the probe.
2. Power-up the Gage by pressing the center navigation        button.
3. Place the probe on the surface to be measured.  When a measurement is calculated,
the Gage BEEPS and the measurement is displayed.

NOTE:

A “---” dashed line value and a low tone indicates the Gage was unable to

calculate the zinc layer thickness 
4. Lift the probe AT LEAST 2 inches (5 cm) between readings or leave the probe on the
surface in the same location for continuous measurements.

Verify accuracy on reference standards at the beginning and the end of each shift, or
if the gage is dropped or suspected of giving erroneous readings.

When enabled (default), the Gage will display two measurement values (as shown
above). To disable, uncheck 

Duplex

within the 

Setup

Menu.  Alternatively, press the

Up

navigation button to toggle Duplex On/Off (when Memory is OFF).

When disabled, the Gage will operate like a conventional combination ferrous/non-
ferrous instrument (similar to the PosiTector 

6000 FNS

probe). Ideal for measuring

non-magnetic coatings over steel and non-conductive coatings over non-ferrous metal
substrates.

N = Paint Thickness 

Zn = Zinc Thickness

Summary of Contents for PosiTector 6000 FNDS

Page 1: ...ither protection system used independently Duplex Mode In Duplex mode the Gage utilizes both magnetic ferrous and eddy current non ferrous principles simultaneously to calculate and display the individual paint and zinc layer thicknesses The magnetic principle is used to measure the combined paint zinc thickness over the ferrous substrate and the eddy current principle is used to measure the paint...

Page 2: ...AT LEAST 2 inches 5 cm between readings or leave the probe on the surface in the same location for continuous measurements Verify accuracy on reference standards at the beginning and the end of each shift or if the gage is dropped or suspected of giving erroneous readings When enabled default the Gage will display two measurement values as shown above To disable uncheck Duplex within the Setup Men...

Page 3: ...nt 1 Select F Zero from the Cal Settings menu 2 Press to select the number of readings to be used to obtain an average typically 3 to 10 readings The greater the variation between readings the more zero readings should be taken to obtain an average 3 Repeatedly measure the uncoated steel part The Gage will wait two seconds after placing the probe on the surface to allow the user to correctly posit...

Page 4: ...o correctly position the probe on the surface After the last measurement the Gage will calculate and display the reading which represents the average of all the measurements taken If the expected reading is not obtained within tolerance lift the probe from the surface and adjust the reading down or up to the expected thickness and press NOTE The reference thickness should be similar to the expecte...

Reviews: