7. Move the sponge over the coated surface at a maximum rate of
0.3 m/s (1 ft/s) and apply sufficient pressure to maintain a wet
surface. If the sponge contacts a pinhole in the coating, all
display LEDs will illuminate and an alarm tone will sound. Alarm
will stop when the sponge is moved away from the pinhole.
8. Precisely locate the pinhole by using a corner of the sponge.
Sponge life depends on use and the abrasive nature of the
surface being tested. Replacement sponges are available with a
comprehensive selection of Accessories (see pg. 8)
PosiTest LPD Control Panel
Buttons:
Power
• Press once to power-up the detector. A self-
test of the 4 test voltages is performed. This
is indicated by a fast flash of each voltage
setting indicator. The self-test should take
approximately 2 seconds to compete. If the
self-test fails, the detector will flash all the
LEDs, sound the alarm then power off.
To conserve battery life, the
PosiTest
LPD
automatically turns OFF after 5 minutes of no
activity.
•
Press and briefly hold
to check the battery
status on the LED display.
•
Press and hold for 3 seconds
to power-down.
Selected
Voltage
and
Audio Alarm
settings
are retained.
The self-test can fail if performed while
the sponge/base tube is in contact with a
grounded surface.
6
NOTE:
NOTE:
NOTE:
Summary of Contents for LPDKITB
Page 1: ...Instruction Manual v 1 0 Low voltage Pinhole Detector LPD ...
Page 15: ...Low voltage Pinhole Detector LPD Manual de Instrucciones v 1 0 ...
Page 29: ...Gebrauchsanweisung v 1 0 Low voltage Pinhole Detector LPD ...
Page 43: ...Notice D Instructions v 1 0 Low voltage Pinhole Detector LPD ...