![Dectris PILATUS3 S Technical Specifications Download Page 18](http://html1.mh-extra.com/html/dectris/pilatus3-s/pilatus3-s_technical-specifications_2462226018.webp)
Information
#5
Briefly pushing the power button on the front panel will shut down the detector control unit. May take up to
1 min.
Figure 4.6:
PILATUS3 S 6M detector control unit as seen from the back.
4.2.2. Connectors
Table 4.5:
Detector Control Unit Connectors
Connector
Description
EM1 (Embedded, figure 4.6)
1 GBase-T adapter
labeled as LAN
User configurable GbE Network Interface
Preconfiguration: DHCP
EM2 (Embedded, figure 4.6)
1 GBase-T adapter
User configurable GbE Network Interface
Preconfiguration:
Static 10.42.41.10 (Netmask 255.255.255.0)
P2P1 (Slot 1, figure 4.6)
Unused
P2P2 (Slot 1, figure 4.6)
10 Gbit SFP+ Ethernet adapter
labeled as PPU (if available)
is reserved for the PILATUS Processing Unit (PPU)
static IP address (10.10.10.10)
The start-up script /etc/rc.local disables ARP. Do not change this!
P3P1 (Slot 2, figure 4.6)
10 Gbit SFTP Ethernet adapter
Detector Interface Port1
labeled as Data1
Static 10.0.10.1
The start-up script /etc/rc.local disables ARP. Do not change this!
P3P2 (Slot 2, figure 4.6)
10 Gbit SFTP Ethernet adapter
Detector Interface Port2
labeled as Data2
Static 10.0.11.1
The start-up script /etc/rc.local disables ARP. Do not change this!
2 x Power (figure 4.6)
AC Connector
See DELL owner’s manual for further details.
Note: ARP is re-enabled if the network service has been restarted. In this case execute the startup script /etc/rc.local as
super user or reboot the system. To assure stable operation of the detector system the configuration file /etc/sysctl.conf is
changed such that the Ethernet rx and tx buffers are larger than the standard setting.
PILATUS3 S 6M Technical Specifications v1.1.1
13 | 25