![Dectris PILATUS3 S Technical Specifications Download Page 13](http://html1.mh-extra.com/html/dectris/pilatus3-s/pilatus3-s_technical-specifications_2462226013.webp)
4. DETECTOR DIMENSIONS AND CONNECTORS
4.1. PILATUS3 S 6M Detector
4.1.1. Technical Drawing
Information
#3
3D step files of the PILATUS3 S 6M detector are available on request. Please contact DECTRIS technical
support for more information.
Rev.
Ausgabedatum
Blatt
Erstellt durch
Kontrolliert von
Dokumentenart: 120402.M.00.L00 Pilatus3 6M Dimensions.idw
Material
Ma stab
Format
F r diese Zeichnung behalten wir uns alle Rechte vor.
Ohne unsere vorherige Zustimmung darf sie weder
kopiert noch Dritten zug nglich gemacht werden.
Allgemeintoleranz
ISO 2768-
DECTRIS LTD, 5400 Baden / Switzerland
A0
1 : 2
erb
120402.M.00.L00
Pilatus3 6M
Dimensions
1 /1
0
06.12.2012
Zeichnung
Rev
Beschreibung/Description
Datum/Date
genehm./approv.
0
Zeichnung erstellt
06.12.2012
erb
228mm Base Plate edge to centre of sensor area
15,64
39,5
5
3
,2
6
0
3
2
5
0
6
0
0
M6x1 - 6H
M6x1 - 6H
M6x1 - 6H
M6x1 - 6H
6,6 Through
6,6 Through
7 Through
7 Through
79,55mm to Sensor
121,55mm to Sensor
138,5mm to front edge without cover
4
5
6
(456)
350
96,5
300
163
58
590
12mm Sensor to mylar foil
17mm Sensor to front edge
155
min.
mm distance from wall
8
9
,5
5
299,7
2
6
3
,3
Center of gravity
Center of gravity
View without cover and Mylar foil
The maximum allowable screw-in depth is 10-16mm
when using M6x1 mounting holes
454,5
0,2
2
6
3
,3
4
0
0
6
2
8
526
5
2
6
4
3
7
,3
4
3
0
8
,2
2
2
18
,6
7
2
6
2
,7
8
424,71
263
Figure 4.1:
Drawing of the PILATUS3 S 6M Detector (also printed separately in the user documentation folder)
4.1.2. Front Side of the Detector
Danger
#1
Danger of electric shock. Do not touch the Mylar
®
foil. The sensors behind the Mylar
®
foil are operated
at high voltages. Touching the Mylar
®
foil can cause an electrical shock.
Warning
#1
Do not touch the Mylar
®
foil to avoid damage of the sensors.
PILATUS3 S 6M Technical Specifications v1.1.1
8 | 25