Change of Main filter:
Remove Pre- and Bag filter
Unscrew fixing screws of Prefilter and Bag filter frame. Remove whole frame.
Loosen nuts of clamping bar and remove clamping bars.
Remove Main filter and discard
Check and clean the sealing surfaces on the unit
Insert the new HEPA filter
Reattach clamping bar
Attach Pre-and bag filter frame and insert filter.
Important:
Main filters must be handled with care. Damages can neutralise the effectiveness
of the filtration.
Hoses and filters are contaminated at first use. Repairs and maintenance work may
only be carried out under compliance with the strictest safety measures.
Fixing screws
Clamping bar
Nut