![DDS Calorimeters CAL3K 3K-F Installation Manual Download Page 14](http://html1.mh-extra.com/html/dds-calorimeters/cal3k-3k-f/cal3k-3k-f_installation-manual_2460696014.webp)
CAL3K-F Installation Guide V5
2022
Website:
www.ddscalorimeters.com
12
Digital Data Systems (Pty) Ltd.
CAL3K Advanced Bomb Calorimeters
www.ddscalorimeters.com
6
DAILY SAFETY STEPS
The 3K-5 vessel is a high-pressure apparatus, and should be treated with care.
•
Handle the vessel over a table or solid surface.
•
Do not drop a vessel.
•
Do not overfill a vessel with oxygen: The maximum oxygen filling pressure is 30bar / 3000kPA, which
is the standard for bomb calorimetry.
•
Do not use any tools to open or close the vessel. If the vessel fails to deflate use the pre-scribed
‘emergency deflate’ procedure as outlined in the CAL3K-F Operations Manual.
•
Use the provided De-filler cap only, part number: 3K-3-22.
•
Do not use any oil or grease on any of the oxygen fittings or the O-rings. A very small amount of high-
pressure (vacuum) grease is allowed on the O-ring to make them pliable. Wipe them clean afterwards.
•
Store the vessel with the lid open. Never store a vessel under pressure, always deflate using the de-
filler cap.
•
Clean the inside of the vessel with paper towel after every use.
•
Weigh the crucible every week and discard it when it has lost 10% of its original mass.
•
Discard the crucible when the bottom is distorted and extremely discoloured.
•
Remove the deflector plate every week and clean the top surface. It does accumulate residue and
must be clean regularly.
•
Make sure to use the pipe-clamping device on the right side of the filing station for safe use.
WATCH YOUTUBE VIDEO ON CAL3K-F
UNPACKING AND INSTALLATION!