- 15 -
HANDPIECE MANUAL MODE
In some situations the POLYLASE MX laser handpiece detection system can be turned
off. This is useful when handpiece-related errors happen in the system and operator needs
to complete the treatment session. In order to start the laser in handpiece Manual Mode,
on the laser controller Start Screen press and hold
(Handpiece Mode) button
for 3 sec. The button status will change from “
Auto
” to “
Manual
” and now the laser can
be started with all handpiece-related errors ignored by the system.
WARNING!
When laser is started in Manual Mode, handpiece spot size can not be
changed as usual or automatically recognized by the laser controller! It is only
possible to work with one currently installed handpiece spot size. If decision was
made to start the laser in Manual Mode, the operator is responsible for the correct
selection of the handpiece spot size (on the laser controller screen) that will match
the actual handpiece output beam spot size. Failure to do so may result in severe
skin burns due to incorrect Fluence settings or other injuries to the patient!
Do not use the laser in handpiece Manual Mode without need and arrange for laser
handpiece service as soon as possible, to eliminate handpiece malfunction that
results in handpiece errors.
SYSTEM SOUNDS
The POLYLASE MX is equipped with the system sounds function. This includes Pulse
Sound and Error Sound options. The laser is factory-configured with Pulse Sound option
turned OFF by default but it could be turned ON from the laser
Settings and Constants
screen in the
Service
Mode
. Error Sound is turned ON by default.
In order to turn system sounds on or off, select
(Service) button from the
Edit
mode, then turn the desired sound option (either both or only one) ON or OFF by the
corresponding slider button in the
Settings and Constants
screen of the
Service
Mode.
Sound ON/OFF Button
can also be used to turn the pulse beep sound ON or OFF
directly from the
Edit
screen
.