Before Beginning
Before you turn on the laser, ensure that the location of the POLYLASE MX Laser does
not interfere with other objects (including the door) in the room. There should be at least
two feet (61 cm) of free air space around all sides of the system.
Treat the optical fiber/handpiece assembly with care: if the handpiece hits the
console, floor or other treatment room equipment, damage can result. Such damage
is not covered in the standard warranty or service contract.
Do not have the main cable stretched so tightly that it puts unnecessary strain on the cable
and its terminations. Be sure that the main cable is out of pedestrian traffic around the
system.
Ensure that protective eyewear appropriate for the Alexandrite / Nd:YAG Laser system
(>OD 7 at 755 nm and 1064 nm) is readily available to be put on
before the system is
energized.
Additional goggles or glasses are available. Please contact Customer Service for more
information.
EVERYONE in the treatment area should have on protective
eyewear at this point, and as long as the system is turned on.