Advanced Channel Emulator
dBm
Operations Manual
ACE600
23
3.
Press
Return
to exit the store utility menu. Note: Register 9 is not
available to store settings as this always contains the factory
defaults.
Figure 2-5. Store display
To re c a ll a s a ve d in s tru m e n t s e ttin g
Pressing the
Recall
key activates the Recall Settings display. A storage
register contains all Instrument State settings, static parameter values,
Update Rate, RF Center frequency, step sizes, and loaded Dynamic file
names. It does not retain the current channel selection, Mode, elapsed
time, or Start/Reset/Pause status.
1.
Press the
Recall
key. The utility display appears. Press the number
of the desired register to be loaded. Recall can be invoked from
Static or Dynamic modes.
2.
Press a number from 1-8, designating a register to recall instrument
settings. Register 0 (labeled
Preset State
) contains the power up
and Preset state. Register 9 (labeled
Default
) contains factory
defaults.
3.
Press
Return
to exit the Recall Settings display. If the recalled
register invokes Dynamic file names that are no longer on the ACE
internal memory, then a "file missing" error is displayed and those
parameters are set to the default value.
Figure 2-6. Recall display
To configure the instrument’s LAN port
1.
Press
LAN
key. The network utility display appears.