MA35-51/E

 

 

NOTE: If the appliance detects an error in the height measurement or the HEIGHT flag is OFF, a screen appears to 
enter the height manually. 

 
 

1.2.1.

 

Weighing children and teenagers 

When users enter an age between 2 and 19 years old, the obtained results of WEIGHT and HEIGHT are shown on a 
chart of percentiles. 
 
Children’s and teenagers’ tickets 

o

 

Children’s and teenagers’ tickets include a table of the most common percentiles, with their reference 
BMI. 

o

 

The table that appears on the ticket corresponds to the child’s sex and age. 

o

 

The values of the tables vary in function of the standard chosen in the appliance setting. 

 
 

1.2.2.

 

Use of the baby scale accessory 

Preparation and precautions 

o

 

Lift up the cradle of the baby scale 

o

 

Before putting the baby in the cradle, press the cradle with your hand to check again that it is securely in 
place 

o

 

While you weigh the baby, watch its movements to make sure that it doesn’t fall from the cradle 

 
Measurement 

The measurement mode for babies has the following features: 
 

o

 

The divisions are by 10 g 

o

 

It only measures weight, not height 

o

 

The appliance puts out a special ticket with information regarding the baby’s weight 

 
To weigh the baby: 
 
(1)

 

Choose the measurement mode “WEIGHT AND HEIGHT” the same as you would with an adult. 

(2)

 

Choose the sex “BABY BOY” or “BABY GIRL” according to the baby 

(3)

 

Enter the age of the baby in months (from 0 to 24) 

(4)

 

Press [START] 

(5)

 

After you put the baby on the cradle, press [START] 

(6)

 

When you finish the measurement, the result screen appears 

 
Baby tickets 

o

 

The baby ticket includes a table of the most common percentiles, with their reference weight. 

o

 

The table that appears on the ticket corresponds to the baby’s sex and age. 

o

 

The values of the baby tables are based on the standards of the World Health Organization. 

 
 

1.2.3.

 

Basal Metabolic Rate (B.M.R.) 

The result of the measurement of weight and height in adult users (20 years or older), can be enlarged with an 
additional screen that includes the calculations corresponding to the Basal Metabolic Rate and the calorie 
consumption recommended in function of lifestyle. 
 
To show this additional screen, make sure that the BMR flag is activated. 
 
 

 

Summary of Contents for cardio prime LCD

Page 1: ...LCD DISTRIBUTOR MANUAL MA35 51 E 22 06 2016 0120 www davicia com ...

Page 2: ... Pressure adjustment 13 2 2 5 Printer adjustments 13 2 2 6 Audio adjustment 13 2 2 1 COMMS RS232 communication adjustment 13 2 3 PROGRAMMING THE MACHINE 14 2 3 1 Programming the clock 14 2 3 2 Programming the language 15 2 3 3 Programming the price 15 2 3 4 Programming the audio 16 2 3 5 Programming the flags 16 2 3 6 Programming the logos 17 2 3 7 Programming the ticket 17 2 3 8 Programming the a...

Page 3: ...r Marcador no definido 6 MANTENIMIENTO Y LIMPIEZA Error Marcador no definido 6 1 PREVENTIVE ACTIVITIES 26 6 2 CLEANING THE DEVICE 26 6 3 CLEANING OF THE BRACELET 26 6 4 CHANGING THE PAPER 26 6 5 APPLIANCE REMOVAL 27 7 RECORDING BASE_14D 28 7 1 LOCATION OF THE FILES 28 7 2 RECORDING PROCESS 28 8 TECHNICAL REVIEW 30 8 1 MODIFICATION OF THE DEVICE 30 8 2 INDICATIONS AND MARKING USE OF GRAPHICS SYMBOL...

Page 4: ...DISTRIBUTOR MANUAL MA35 51 E 3 ...

Page 5: ...creen 4 Loudspeaker 5 Body fat sensors 6 Vascular sensor 7 Barcode reader 8 Coin mechanism 9 Wrist bracelet 10 Acess to the printer and coin drawer 11 Ticket exit 12 Baby scale atachment option 13 Technical service access 14 Switch and power 15 Wheels 1 2 3 4 5 6 7 8 9 10 11 12 13 15 14 4 ...

Page 6: ...sexposed to vibrations or close to heat sources WARNING Do not position the equipment so that it is difficult to disconnect the main switch WALL 2 2 STARTING UP THE APPLIANCE 1 Connect the device to mains using the mains lead The outlet must be grounded Ungrounded adapters should not be used DANGER To avoid the risk of electric shock this equipment must be connected to a mains supply with protecti...

Page 7: ...e warning NO PAPER This indicates that the printer doesn t have paper and can t print result tickets The warning disappears automatically when the sensor detects paper 1 1 2 Sex and age As part of the measurement process you will be asked to select your sex and enter your age This information is especially relevant in the following cases o Calculation of the results of the Body Fat Index o Weighin...

Page 8: ...ement mode for babies has the following features o The divisions are by 10 g o It only measures weight not height o The appliance puts out a special ticket with information regarding the baby s weight To weigh the baby 1 Choose the measurement mode WEIGHT AND HEIGHT the same as you would with an adult 2 Choose the sex BABY BOY or BABY GIRL according to the baby 3 Enter the age of the baby inmonths...

Page 9: ...ended o Just after energetic physical exercise o After a sauna o After a bath o After drinking alcoholic beverages o After drinking a lot of water or after eating it s a good idea to wait 1 or 2 hours o During menstruation Precautions Although no side effects are known the Body Fat Index shouldn t be used in the following cases o During an acute contagious illness o On people with medical implants...

Page 10: ... warning beforethe measurement 1 3 1 Body Composition The result of the measurement of the Fat Index can be extended with an additional screen that includes the calculation of Body Composition That screen shows the results of Weight and Fat along with the estimated amounts of Water Mineral and Protein which correspond to the user s Non Fat Mass To show this additional screen make sure that the W P...

Page 11: ...it is necessary to put the index finger of your left hand on the sensor as indicated on the screen o It is very important to put your finger completely inside the sensor Otherwise the measurement might not be precise o To start the measurement press the START button Make sure that when you press START you have the index finger of your left hand inside the sensor Otherwise the measurement won t be ...

Page 12: ...djustment of the printing density AUDIO Adjustment for the voice volume COMMS Adjustment for the communications port usage 2 2 1 Weight adjustment To adjust the weight follow these steps 1 Make sure that there is no weight on the platform and press any button 2 Place a pattern mass on the platform and press any button 3 Enter the pattern mass value you have placed in grams For example for a value ...

Page 13: ...sted correctly Nevertheless if VREF or BF appear with an NOK when you access the body fat adjustment follow these steps VREF adjustment 1 Press ADJUSTMENT VREF 2 The appliance automatically adjusts the tension tomeasure the body fat and returns to the setting menu BF setting 1 Press ADJSUTMENT BF 2 Disconnect the Body Fat wire from the BASE_14 board and enter the resistance that matches the value ...

Page 14: ...ed The selection of the printer is made using a C028 connector located in the BASE_14 board With connector model TS1700 SEIKO Without connector model MT102 EPSON TS1700 adjustment If the selected model is TS1700 the display shows a message with its reference MT102 adjustment If the model selected is MT102 the version must be set 1 Select the corresponding version according to the reference indicat...

Page 15: ...the clock LANGUAGE Programming the language PRICES Programming the prices AUDIO Programming the volume of the audio FLAGS Programming the flags LOGOS Programming the logos TICKET Programming the ticket ADVERTISING Programming the publicity CUSTOMER Programming the client options ACCOUNTING Programming the accounting HEIGHT Programming the height PRESSURE Programming the pressure diastolic GSM Prog...

Page 16: ...e price of the measurements 1 Press COINS 2 Press the button of the measurement you want 3 Through the keyboard enter the desired amount for the selected measure o During the entering of the price it is not necessary to enter decimal or currency symbols The type of coin as well as the number of decimals of the price is established according to the settings o Press STOP if you want to modify the in...

Page 17: ...te the Body Fat Index measurement W P M Activate or deactivate additional information about Body Composition ATHLETIC FAT ON The appliance will ask for the build of the user Normal or Athletic to determine the result of the Body Fat Index measurement OFF The appliance always operates in Normal mode SELECT POPULATION Select the type of population in order to determine the result of the Body Fat Ind...

Page 18: ...2 Press the keys 1 and 5 to view the available logotype options 3 Press START to confirm any option The selected logotype will appear in the main screen as well as in the printed ticket header as long as the corresponding flags are activated 4 Press STOP to finish 2 3 7 Programming the ticket The available options with this menu are the following PROGRAMMING Programming the ticket lines CODIGO BAR...

Page 19: ... Programming the advertising With this option you can programm a text which will appear in the display distributed in 4 lines It is possible to programm a total of 17 characters on each line taking into account that white spaces are considerated as a character The appliance will center the text automatically 1 Select ADVERTISING 2 Select option 1 3 Enter the information that will appear in the sel...

Page 20: ... the pressure With this option we can modify the value in the diastolic pressure result of the appliance applying the desired percentage on the result 1 Press the keys 1 and 5 to move through the available options 2 Press START to confirm an option 3 Press STOP to exit 2 3 13 Programming the GSM The available options in the menu are the following PIN Programming the PIN code of the phone card whic...

Page 21: ...m 000 to 999 of the amount of printed tickets o When the appliance reaches the settled quantity a sms will be sent with the serial number of the appliance and the corresponding warning o When introducing the value 000 the appliance will not send any message Warning No use 1 Select the option NO USERS 2 Enter the quantity from 000 to 999 of the amount of hours the appliance do not receive any coin ...

Page 22: ...onfirm the information IMPORTANT If we change the BASE_14 board it is very important to reprogram the serial number 2 4 TEST The options to which you have access via the Test menuare as follows WEIGHT Weight test HEIGHT Height test BODYFAT Resistance test PRESSURE Static pressure test BARCODE Barcode reader test VASCULAR HEALTH Communication test with the vascular health sensor board RS232 RS232 T...

Page 23: ... manometer and check that the pressure value matches the appliance s reading 5 1 4 Barcode test Scan different barcodes with the reader and verify that the number displayed matches the number scanned 5 1 5 Vascular health test This test automatically checks the communication with the vascular health sensor board If it is correct it will show OK Otherwise NOK 5 1 6 RS232 test To do the RS232 test i...

Page 24: ...heck 5 On the next screen the appliance will automatically activate the selected item 6 Check that it works properly NOTE It is possiblethat not all outlets will be operational 5 1 13 Input tests This test is used to check that the inputs work 5 1 14 Pressure cycle test The appliance will test the pneumatic appliance compressor electrovalve and bracelet through continuous cycles of pressure and de...

Page 25: ... to test the communication channel RS232 and the reception value Depending on the signal value it shows the coverage 3 Press STOP to exit Communications 1 Select the option COMMS 2 The appliance sends a test message to the sttled phone see Programming GSM 3 Press STOP to exit 5 1 22 Pressure switch test This test checks the pressure switch state which stops the bracelet inflation when it reaches t...

Page 26: ...rd failure 40 VASCULAR HEALTH Vascular board failure Vascular sensor failure Failure in the wiring between the sensor and the Vascular board Failure in the main wiring of the Vascular board 41 VASCULAR COMMUNICATIONS Vascular board failure Failure in the main wiring of the Vascular board 70 BASE_14 Unknown failure in the BASE_14 board 71 BASE_14 E2PROM 24C01 PRESS V3 Failure in the E2PROM 24C01 of...

Page 27: ...thorized Technical Assistance Service DAVI CIA personnel 6 2 CLEANING THE DEVICE o In order to clean the appliance turn it off and disconnectthe mains lead o Clean the body of the appliance using only a slightly damp soft cloth o Do not use gasoline or any similar solvent product 6 3 CLEANING OF THE BRACELET o Clean the bracelet using a cloth soaked in a mixture of water and a neutral detergent o ...

Page 28: ...se the front door 6 5 APPLIANCE REMOVAL Never throw the product away with domestic waste When you want to get rid of it take it to a proper dumping site If you don t know where the closest one is ask your city hall The wrapping of this product is recyclable place it in a container for paper and cardboard The symbol of the container with the cross means that when the appliance is no longer useful i...

Page 29: ...4D The files in BIN format should be located in the Flash folder inside the DAVIFlash program directory prior to starting the recording process IMPORTANT The recording files have a suffix and an extension _1 bin or _2 bin corresponding to the number of the flashthat is recorded Don t change the name of the files as this will prevent the recording process 7 2 RECORDING PROCESS 1 Turn off the scale ...

Page 30: ...s STOP and select the option 1 TEST COMMS again c If there is no communication after 3 tries check the connections 10 Once the communication test has been performed successfully select the option 2 PROGRAM FLASH 11 Select the option that corresponds to the flash that youwant to record 12 Press START to start the recording The recording process lasts about 10 minutes per flash 13 When the recording...

Page 31: ... from the manufacturer WARNING If you modify this equipment you must perform adequate inspections and tests to ensure the continued safe use of the equipment For parts or components that can be replaced DAVI CIA will provide circuit diagrams component lists descriptions calibration instructions or other information 8 2 INDICATIONS AND MARKING USE OF GRAPHICS SYMBOLS 7 2 1 Marketing equipment Graph...

Page 32: ...ymbol Description Storage position Fragile Keep out of reach of the rain 7 2 3 Software symbols Graphic symbol Description Do not use if you are pregnant Do not use if you have a pacemaker Do not use the clock since 7 3 MOVING THE DEVICE WARNING Slide the device from the back 31 ...

Page 33: ...ment you should stand straight and not move 7 4 2 Body fat index The patient holds the side handles of the team ensuring that the palm is in contact with the two electrodes of each handle Thus the equipment can measure the body impedance and the estimation Body fat index 7 4 3 Blood Pressure The patient introduces the wrist of the left arm in the air cuff After pressing START wait to complete the ...

Page 34: ...ft hand in the finger sensor vascular and then pressing START wait to complete the measurement During this process you should not speak or move 7 5 MATERIALS Component Material Risk Case Polyethylene No risk Handles Aluminum No risk pneumatic bracelet polyamide No risk Vascular sensor Silicone No risk 33 ...

Page 35: ...Pressure Fat index Vascular health 300 kg load cell Ultrasound Oscillometric method Bioelectric Impedance Analysis BIA Fotopletismografy Accuracy of the measurements Weight Height Static pressure Fat index Vascular health Non Metrologic 100 g n 3000 Metrologic C accuracy class n 3000 1 cm 3 mmHg 2 2 Power supply 230V 50 60 Hz 115V 50 60 Hz Consumo 40 VA máx Environmental conditions Operation Fat i...

Page 36: ...e T 2 x 0 800 A 5 x 20 mm Delayed Type T Classification according to electrical safety standard EN 60601 1 Continuous operation Bioelectrical Impedance Analyzer Type BF Photoplethysmograph Type BF Surge protection Class I Protection against ingress of water Ordinary Not suitable for use in presence of flammable anesthetic mixture with air oxygen or nitrous oxide 35 ...

Page 37: ...t 17 Specific conditions for Broadband Data Transmission Systems ELECTROMAGNETIC EMISSION EN 55011 2009 A1 2010 Radiated Class B EN 55022 2010 AC 2011 Radiated Class B EN 55011 2009 A1 2010 Conducted Class B EN 55022 2010 AC 2011 Conducted Class B EN 55014 2006 A2 2011 Discontinuous Conducted Class B EN 61000 3 2 2006 A1 2009 A2 2009 Harmonics Class A EN 61000 3 3 2013 Voltage fluctuations and fli...

Page 38: ... of charge upon presentation of the purchase receipt for the appliance No other rights can be taken into consideration Return shipment is at the customer s expense if the appliance is anywhere other than at the customer s headquarters In the event of damage during shipment guarantee rights will be granted only if the appliance is shipped in the original packaging You are therefore recommended to k...

Page 39: ... supporting documents are available on the website www davicia com C Murcia 35 Nave G Pol industrial Can Calderón E 08830 Sant Boi de Llobregat Barcelona Spain Tel 34 93 654 32 32 Fax 34 93 654 16 62 info davicia com www davicia com 38 ...

Reviews: