USER’S MANUAL
MA18-02/B
13
6. TROUBLESHOOTING
Error Message
Error Message
Error Message
Error Message
Reason
Reason
Reason
Reason
Action
Action
Action
Action
Low Battery: This warning shows
that the voltage of battery is too
low to use
Please replace a new battery or plug the AC
adaptor for operation
Overload: The total load exceeds
the maximum capacity of the scale
Please reduce the loading and try again
Counting error (too high): Indicates
that the signal from the load cell is
too high
This error is normally caused by a serious fault on
the scales such as faulty load cell or wiring. Please
contact the local service representative.
Counting error (too low): Indicates
that the signal from the load cell is
too low
This error is normally caused by a serious fault on
the scales such as faulty load cell or wiring. Please
contact the local service representative.
Zero count over calibration zero
range +10% while power on
Please recalibrate the instrument
Zero count under calibration zero
range -10% while power on
Please recalibrate the instrument
EEPROM Error: Indicates that
there is a fault with the scale’s
software
This error is normally caused by a serious fault on
the scales such as faulty load cell or wiring. Please
contact the local service representative.