Page 4
Tec 5 Vaporizer
August 1999
O & M Manual Part No. 1105-0100-000
TEC 5
SEVOFLURANE
w
Warning: Earlier versions of the Selectatec Series Mounted Manifold which provide mounting
positions for three vaporizers require that if only two vaporizers are fitted then the centre position
must be occupied. If the centre position is not occupied the interlock which helps to ensure that only
one vaporizer at a time can be turned ON is ineffective.
Later versions of the Selectatec Series Mounted Manifold which provide mounting positions for three
vaporizers incorporate an additional interlock which helps to ensure that only one vaporizer at a time
can be turned ON even if the centre position is not occupied.
3.4
Vaporizer Identification Label
A vaporizer identification label is affixed
to the back panel of the vaporizer as
illustrated on Fig. 1.
An anaesthesia system fitted with a
vaporizer identification unit uses this label
to identify the vaporizer type.
Do not affix any additional labels or
markings to the back panel, they may
adversely affect the operation of the
vaporizer identification unit.
Fig. 1 Vaporizer Identification Label
4.0
Operating Instructions
4.1
General
The vaporizer is designed to be used on Selectatec Series Mounted Manifolds. The vaporizer can be
installed on other Selectatec Manifolds but the interlock system is designed to function on Selectatec
Series Mounted Manifolds only.
Mounting a Tec 5 vaporizer on a Selectatec 7 Compatibility Block is not recommended.
w
Warning: It is a requirement of European Standard prEN 740 - Anaesthetic Workstations And Their
Modules, that the gas monitoring device referred to at 3.1 of this manual is in operating condition,
by being enabled and functioning , prior to use of the vaporizer.
w
Warning: Handle the vaporizer with care at all times. Do not lift or support the vaporizer by holding
the control dial.
w
Warning: Before mounting a vaporizer on to the manifold ensure that only one 'O' ring is fitted to
each manifold port valve. Ensure that each 'O' ring is intact and that there is no foreign matter
around the mating surfaces. A damaged 'O' ring and/or foreign matter around the mating surfaces
can cause leaks.
w
Warning: Do not use a vaporizer if the liquid level decreases to below the minimum level.
w
Warning: Before using a vaporizer allow it to attain the ambient temperature of the location in
which it is to be used.
IDENTIFICATION
CODE
AGENT NAME
VAPORIZER
NAME TEC 5
VAPORIZER
IDENTIFICATION
LABEL
AA.13.051