Tec 5 Vaporizer
Page 5
O & M Manual Part No. 1105-0100-000
August 1999
1
2
3
PUSH DOWN
TO FULL EXTENT
4.2
Vaporizer Mounting Procedure
1.
Remove the red coloured dust caps fitted to the vaporizer interlock block ports, illustrated on Fig. 2.
2.
Check that each port valve 'O' ring is intact and not damaged. Renew each 'O' ring if necessary.
3.
Make sure that the vaporizer control dial is in the OFF position.
4.
Make sure that the vaporizer locking lever is at the unlocked position, that is, in the up position.
Fig. 2 Vaporizer Mounting Procedure
5.
Carefully lower the vaporizer on to the manifold so that the vaporizer interlock block covers the two
manifold port valves with the interlock block ports correctly engaged with the manifold port valves.
w
Caution: Push the locking lever down to the full extent of its travel before attempting to turn it. The
mechanism may be damaged if an attempt is made to turn the lever before it is at the full extent of
its vertical travel.
6.
Press down the locking lever
1
to the full extent of its travel and then turn it clockwise
2
to the
locked position
3
to lock the vaporizer on to the manifold as illustrated on Fig. 3.