Datex-Ohmeda S/5 Anesthesia Monitor
A-2(3)
Document No. 800 1005-1
16.
Flow rates
Sampling flow
180-220 ml/min
Side flow
10-27 ml/min
O
2
meas. in flow
45-60 ml/min
O
2
ref. in flow
25-42 ml/min
CO
2
absorber flow
180-220 ml/min
17.
Pressures
Amb-Work
50-75 mmHg
OM (in-ref)
> 4 mmHg
18.
O
2
sensor output voltage
min. 2800 mV
OK
N.A.
Fail
19.
Gas calibration
20.
Fall time measurement
CO
2
fall time
< 360 ms
O
2
fall time
< 480 ms
N
2
O fall time
< 360 ms
AA fall time
< 520 ms
21.
Noise measurement
CO
2
noise
< 20
O
2
noise
< 100
N
2
O noise
< 150
AA noise
< 20
AA ID option
22.
ACX_ASX Delay
400-800 ms
23.
Peak normal and Peak mirror
Peak normal
10.30-10.70
Difference
±
0.30