S/5 Airway modules
Document No. 800 1005-1
13
Figure 6
Gas sampling system layout
See new sampling system layout since autumn 1998 in figure 8.
Filter
CO
2
absorber
D-fend
ACX-200
Zero valve
Pressure valve
O
2
sensor
ASX-200
Damping
chamber
Sample pump
Ref IN
Room AIR
Sample
Gas IN
Pressure
transducer
Sample
Gas OUT
A
B
C
side flow
main flow
Constriction cassette
in G-AiO/AiOV
Figure 7
Gas sampling system block diagram
In G-AO, -AOV models, tube A is Teflon, B and C Nafion. In G-AiO, -AiOV models, tubes A and C are
Nafion, B is Teflon. Figure 7 is valid for modules manufactured since autumn 1998.