Datex-Ohmeda S/5 Anesthesia Monitor
Document No. 800 1005-1
58
PVX rep status
FFFF continuously
General Status
0
Normal operation state
8000
Initialization
If not 0 or 8000, replace the gas mother board.
Timeouts
is a cumulative number that indicates how many times the module has not responded to
the monitor's inquiry.
Bad checksums
is a cumulative number that indicates how many times communication from the
module to monitor broke down.
Bad c-s by mod
is a cumulative number that indicates how many communication errors the
module has detected.
The monitor starts counting these items at power up and resets to zero at power off. The nonzero
values do not indicate a failure, but the continuous counting indicates either serial communication
failure or module not in place.
The timeouts etc numbers should not grow faster than 50/s.