3 Operation and Tutorial
1009-0307-000
3-19
PEEP
PEEP is only available during mechanical ventilation. But, you can set the
control at any time.
The range is off or between 4 and 30 cm H
2
O.
Note: When adjusting PEEP, a digital value is displayed to assist PEEP setting
until the setting is confirmed. The PEEP pressure value in the monitoring
section of the display will indicate the actual PEEP pressure being delivered.
No value is shown if PEEP is off.
w
W
W
W
WAA
A
ARR
R
RN
N
N
NIIIIN
N
N
NG
G
G
G
Do not use a separate mechanical PEEP valve; incorrect operation and
patient injury can result.
Set inspiratory pause
You can only use inspiratory pause in volume control.
When Pause is on, the inspiratory volume stays in the patient’s lungs for the
set pause time at the end of inspiration.
Pause can be set to off or 5 to 60 percent of inspiratory time in increments of
five percent.
When the ventilator is turned on, the inspiratory pause is at the value set when
the ventilator was last used.
or
30
Step 1
Push the Menu key to see
the Main Menu.