S/5 Anesthesia Monitor and S/5 Critical Care Monitor
Document No. 8002948
14
4.1.2
Digital section
Digital section faulty.
Remove all modules and boards
except CPU, UPI and display
controller board
OK ?
A
beep is
heard when
monitor is
turned
on ?
Insert one of the removed
boards.
Turn
the monitor on.
OK ?
Replace faulty board
Replace CPU board.
Is tone
same as at
OK start but
longer ?
Is tone
lower and longer
than at OK
start ?
Replace Display controller.
Replace UPI-board
Yes
No
No
No
No
Yes
Yes
Yes
Yes
Is tone
same as at
OK start ?
Try with another display
OK?
Yes
Yes
No
No
Replace display
Turn
the monitor on.
Figure 4
Digital section troubleshooting flowchart