S/5 Anesthesia Monitor and S/5 Critical Care Monitor
Document No. 8002948
10
14.
Perform an electrical safety check and leakage current test.
?
15.
Check that the Central Unit functions normally after the performed electrical safety check.
?
•
Fill in all necessary documents.
3.3
Disassembly and reassembly
NOTE: Switch the monitor to standby and press the service reset -switch at least for five seconds
before detaching any PC boards.
1.
Remove screws and detach all PC boards and cover plates from the right side of the CPU
board.
2.
Remove the screws on the CPU board.
3.
Detach the CPU board.
Reassembly should be made in reversed order.
NOTE: When reinstalling PC boards, push the boards carefully until they stop before fastening them
with screws.
3.3.1
Downloading/replacing software on CPU Board, B-CPU5/B-CPU4
See also
“Replacing CPU Board, B-CPU5/B-CPU4”
on
Part I/System Installation
.
NOTE: All user settings will be lost after downloading/replacing monitor software.
NOTE: During the downloading of monitor software the serial number of the CPU board is written
onto the software card and is matched with the downloaded monitor software serial number. If the
downloading of monitor software would fail, the same monitor software could be downloaded
again only onto the same CPU board, but not onto any other CPU board.
1.
Make sure that the monitor is turned to standby. Press and hold the service reset switch on
the Central Unit rear panel for at least five seconds, or until an audible tone is generated.
2.
Move off the lid for software card on the B-CPU5 / B-CPU4 rear panel.