![Datex-Ohmeda ARK N-SCAN Technical Reference Manual Slot Download Page 20](http://html.mh-extra.com/html/datex-ohmeda/ark-n-scan/ark-n-scan_technical-reference-manual-slot_513827020.webp)
Datex-Ohmeda S/5 monitors
Document No. 8001019-3
16
5.1.1
Keyboard Log
All the keyboard presses and the commands given by the
ComWheel are recorded in the Keyboard Log. The
keyboard log is saved in the permanent memory of the
monitor. The length of the log is 1150 events. The log is
FIFO type.
5.1.2
Keyboard Type
Store Mask
A selection for setting the Keyboard for the
deioRecorder for Anesthesia’s language. The selected
language determines the outcome of the lower keypad.
Store Type
is for setting the keyboard’s type;
COM = Command Board
ARK = Anesthesia Record Keeping Keyboard
IC = Information Center Keyboard
NOTE: The settings should be checked if the
controller board is replaced. If settings are
changed, the new settings will not be valid until
the next start-up.
Summary of Contents for ARK N-SCAN