![Datex-Ohmeda ARK N-SCAN Technical Reference Manual Slot Download Page 14](http://html.mh-extra.com/html/datex-ohmeda/ark-n-scan/ark-n-scan_technical-reference-manual-slot_513827014.webp)
Datex-Ohmeda S/5 monitors
Document No. 8001019-3
10
3.
Check external parts:
−
the keyboard for the deioRecorder for Anesthesia, K-ARKB plastic frame is intact
−
the front panel stickers are intact
−
the ComWheel cover is intact and is attached properly
−
all four rubber pads are in place on the bottom cover
?
•
Install the B-ARK into the Central Unit. Connect and lock the interface cable to the B-ARK
rear panel connector.
4.
Switch the monitor on. Check that the LED on the upper right hand corner of the keyboard
for the deioRecorder for Anesthesia, K-ARKB is lit up.
?
5.
Enter the service menu:
Monitor Setup
- Install/Service
(password 16-4-34)
- Service
(password 26-23-8)
Take down the information regarding keyboard for the deioRecorder for Anesthesia, K-ARKB
software.
?
6.
Select the menu
Keyboard
with the ComWheel.
Highlight the text
Upper Led
. Check that the red alarm LED is turning on and off on the
keyboard for the deioRecorder for Anesthesia when pressing the ComWheel. Check also
the yellow alarm LED by selecting
Lower Led
from the menu.
?
7.
Check the ComWheel.
Turn the ComWheel clockwise and counterclockwise and check that each step generates a
sound from the loudspeaker and the corresponding values at the bottom of the menu
increase.
Select
Dummy Press
. Press the ComWheel and check that the press generates a sound and
the corresponding value in the menu increases.
?
8.
Check the keyboard for the deioRecorder for Anesthesia, K-ARKB membrane keys.
Press the keys on the upper part of the Keyboard for the deioRecorder for Anesthesia one by
one. Check that each key generates a sound from the loudspeaker and the corresponding
text in the menu changes from yellow to red.
Summary of Contents for ARK N-SCAN
Page 2: ......
Page 4: ...Datex Ohmeda S 5 monitors Document No 8001019 3 ii SERVICE CHECK FORM A 1 ...
Page 26: ...Datex Ohmeda S 5 monitors Document No 8001019 3 22 ...
Page 27: ...APPENDIX A Service check form K ARKB Document No 8001019 3 23 APPENDIX A ...
Page 28: ...Datex Ohmeda S 5 monitors Document No 8001019 3 24 ...
Page 30: ...Datex Ohmeda S 5 monitors Document No 8001019 3 ...