Datex-Ohmeda S/5 monitors
Document No. 8001019-3
12
4
TROUBLESHOOTING
4.1
Troubleshooting charts
4.1.1
Keyboard for the deioRecorder for Anesthesia, K-ARKB
Problem
Cause
Treatment
Keys have not effect on the display.
Cable is not connected or broken.
Wrong type of cable is connected.
Loose connector inside.
Component failure inside.
Connect right type of cable properly (see above).
Detach the bottom plate and check connectors
and components.
Membrane key not working.
Ribbon cable loose or broken. Keyboard cable
loose or broken. D-connector pin failure. IC
failure on the Controller board. RS232
communication failure on the main CPU board.
NOTE: The cancel key does not respond if the
menu is closed.
The modify key may not work if there is no
selection.
Check the items. Replace them if necessary.
Led does not light at alarm or stays
lit after alarm is over.
Cable loose or broken. LED broken.
Component failure on the Controller board.
Check the items. Replace them if necessary.
See more troubleshooting items on
User’s Reference Manual
.
Summary of Contents for ARK N-SCAN
Page 2: ......
Page 4: ...Datex Ohmeda S 5 monitors Document No 8001019 3 ii SERVICE CHECK FORM A 1 ...
Page 26: ...Datex Ohmeda S 5 monitors Document No 8001019 3 22 ...
Page 27: ...APPENDIX A Service check form K ARKB Document No 8001019 3 23 APPENDIX A ...
Page 28: ...Datex Ohmeda S 5 monitors Document No 8001019 3 24 ...
Page 30: ...Datex Ohmeda S 5 monitors Document No 8001019 3 ...