Trying Out the Instrument
This chapter describes some basic functions that you can perform with the
R&S
NGL/NGM.
Source and sink current
The R&S
NGL/NGM power supply series are 2 quadrant power supplies
which may both source and sink current. When the voltage across the out-
put terminal exceeds the set voltage, current flows into the instrument. The
default behavior "Auto" can be configured in output menu, see section "Out-
put mode" in the User Manual.
On the display, sink mode is shown as negative current. See also
5.1
Setting the Output Voltage and Current
1. Press [Home] key.
The R&S
NGL/NGM displays the home window.
2. Select voltage or current parameter in the home window.
The R&S
NGL/NGM displays an on-screen keypad to set the value.
3. Enter the required value.
4. Confirm value with the unit key (V/mV or A/mA).
on page 24 for more information on the operat-
ing modes supported in different models.
5.2
Activating the Channels Output
The output voltages can be switched on or off regardless of the instrument's oper-
ating mode.
Activating the Channels Output