Cleaning the counting detectors
From time to time, dust may be stuck on the counting detectors.
This can cause miscounts.
When a counting process is initiated and the machine detects an
obstruction on one of its detectors it will show an error message of
“Clean detector 1” or “Clean detector 2”. In some of the cases it
can also show “Call Service Er1” or “Call Service Er2”.
A special cleaning mechanism was designed for the SEED count
JR. To clean the detectors properly, follow the next steps:
1. Turn off the machine and unplug it from the electricity.
2. Rotate the upper case of the machine counterclockwise
one click, until you see the air nozzle on the left side (see
picture).