9.
For the DT9837A-OEM and DT9837B modules, you can also measure the time between
the A/D sample to the specified edge of the gate input signal, the time between two gate
input signals, or the time between the specified edge of the gate input signal to the A/D
sample by specifying channel 6 (gate counter 2) in the analog input channel list; see
for more information. Configure the Gate.CT2 measurement parameters as follows:
a.
From the
Start Select
combo box, select the signal that you want to use to start the
measurement (
A/D Sample
,
Gate Rising Edge
, or
Gate Falling Edge
).
b.
From the
Stop Select
combo box, select the signal that you want to use to stop the
measurement (
A/D Sample
,
Gate Rising Edge
, or
Gate Falling Edge
).
Note that if you choose to start the measurement using the A/D sample, choose a gate
edge to stop the measurement. Likewise, if you choose to stop the measurement using
the A/D sample, choose a gate edge to start the measurement.
c.
From the
Self Clear
combo box, select
Clear to zero
if you want to return a value of 0
between measurements, or select
Keep last Value
if you want to return the previous
measurement value between readings.
10.
For the Coupling type, select
AC
for AC coupling or
DC
for DC coupling.
11.
For the Current Source, select
Enabled
to enable the internal excitation current source or
Disabled
to disable the internal excitation current source.
Note:
If you enable the use of the internal excitation current source, it is recommended
that you choose AC coupling. Refer to
for more information on wiring IEPE
inputs.
12.
When you are finished, click
OK
to close the Measurement Options dialog box.
13.
Repeat steps 4 to 11 for the other modules that you want to configure.
14.
When you are finished configuring the modules, click
Close
to close the Control Panel.
Note:
If you need to configure these settings programmatically instead of using the Open
Layers Control Panel, you can write to the registers of the module. Refer to
for more information.
Summary of Contents for DT9837
Page 1: ...DT9837 Series UM 22417 AJ User s Manual Title Page ...
Page 4: ......
Page 12: ...About this Manual 12 ...
Page 26: ...Chapter 1 26 ...
Page 27: ...Part 1 Getting Started ...
Page 28: ......
Page 53: ...Verifying the Operation of a Module 53 ...
Page 60: ...Chapter 4 60 ...
Page 61: ...Part 2 Using Your Module ...
Page 62: ......
Page 90: ...Chapter 5 90 ...
Page 104: ...Chapter 6 104 ...
Page 110: ...Chapter 7 110 ...
Page 130: ...Appendix A 130 ...
Page 144: ...Appendix C 144 ...
Page 150: ...Index 150 ...