Chapter 7
84
Calibrating the Voltage Offset and Gain
To calibrate the 24-bit A/D offset and A/D gain circuit on the DT9828 module, you must use a
precision voltage standard, such as the Krohn-Hite Model 523 or equivalent, and a shorting
plug.
To calibrate the voltage offset and gain, perform the following procedure:
1.
Wire a shorting plug to analog input channel 0.
2.
From the Voltage Calibration section of the dialog box, click
Run.
3.
Ensure that 0 V is applied to analog input channel 0, and then click
OK
.
The voltage offset value is calibrated automatically.
4.
Apply 140 mV to analog input channel 0, and then click
OK
.
The voltage gain value is calibrated automatically.
Note:
If you are not satisfied with the analog input calibration, you can load the factory
default settings stored in the EEPROM by clicking
Restore Factory Settings
.
Summary of Contents for DT9828
Page 1: ...DT9828 User s Manual UM 24984 A Title Page ...
Page 4: ......
Page 8: ...Contents 8 ...
Page 12: ...About this Manual 12 ...
Page 13: ...13 1 Overview Features 14 Supported Software 16 Accessories 18 Getting Started Procedure 19 ...
Page 20: ...Chapter 1 20 ...
Page 21: ...Part 1 Getting Started ...
Page 22: ......
Page 30: ...Chapter 2 30 ...
Page 49: ...Verifying the Operation of a Module 49 ...
Page 54: ...Chapter 4 54 ...
Page 55: ...Part 2 Using Your Module ...
Page 56: ......
Page 57: ...57 5 Principles of Operation Analog Input Features 59 Digital I O Features 67 ...
Page 80: ...Chapter 6 80 ...
Page 86: ...Chapter 7 86 ...
Page 92: ...Chapter 8 92 ...
Page 106: ...Appendix A 106 ...
Page 112: ...Appendix B 112 ...