DARAY V402+ Operating Manual Download Page 35

Summary of Contents for V402+

Page 1: ...V402 HANDHELD PULSE OXIMETER TEMP MONITOR OPERATING MANUAL QAM V402 1 02 07...

Page 2: ......

Page 3: ...NS 6 3 INSTALLATION 8 4 DISPLAY AND OPERATION 9 5 SPO2 MONITORING 18 6 TEMP MONITORING 22 7 MAINTENANCE 23 8 V402 PATIENT MONITOR DATA VIEWER 26 9 PACKING LIST AND ACCESSORIES 26 RETURNS POLICY 29 WAR...

Page 4: ...f the relevant room complies with the applicable national and local requirements This product is operated under strict observance of this manual The safety statements presented in this chapter refer t...

Page 5: ...xplosive substances vapours or liquids Do not pull or lift the monitor by its cables It could lead to falling and consequent patient injury It is not recommended to hang the monitor when transporting...

Page 6: ...ulse rate PR plethysmogram waveform PLETH pulse strength and temperature TEMP The monitor measures these parameters through a SPO2 sensor and a TEMP sensor and displays them on the LCD screen The moni...

Page 7: ...eft in setup or viewing trend graphs 8 AC power adapter socket 9 PC serial link connector socket 10 Up arrow backlight Turns backlight on off whilst monitoring or moves cursor up in setup or viewing t...

Page 8: ...adaptor s power indicator LED lights up Insert the adaptor s connector to the adaptor socket on the monitor s right side as shown in Fig 1 8 WARNING The AC power adaptor supplied is built to special...

Page 9: ...removed from the sensor the monitor automatically enters standby mode In this mode when the SPO2 sensor is reconnected and a finger is inserted into the sensor or the TEMP cable is reconnected the mon...

Page 10: ...only the SPO2 sensor is connected to monitor a patient s SPO2 the LCD screen displays the measurements of SPO2 PR and pulse strength status information and system information see Fig 3 Fig 3 TEMP Scre...

Page 11: ...en displays PLETH waveform and measurements of the parameters See Fig 6 Fig 6 No Title Description 1 SPO2 SPO2 value refresh 1 second 2 Alarm audio ON OFF 3 Pulse strength Max 10 segments 4 Pulse audi...

Page 12: ...screen If the speaker sound is turned off by pressing the Mute button or the parameter alarm switch is set off the parameter alarm sound will be silenced but the parameter s measurement still flashes...

Page 13: ...cription of Information Displayed The meanings of the words and the ranges of the values in Fig 7 are listed in table 4 2 as follows Upper limit Lower limit Alarm SPO2 85 100 85 100 ON or OFF PR 40 25...

Page 14: ...the maximum and start decreasing again With the cursor on the alarm on off positions press the up button to toggle between ON and OFF When the Down arrow button is pressed with the cursor on the DATA...

Page 15: ...ue to the menu button is pressed on the other screen the monitor draws the parameter trend waveforms on the time measurement coordinate series One screen can only display 3 hours data which is defined...

Page 16: ...ator The data forming one screen trend waveforms time period of 3 hours is defined as a page which corresponding to 90 units of the storage area The page indicator points out the position of this data...

Page 17: ...or left and right in single steps but remains within the current page of data Up and Down Arrow Buttons The up and down arrow buttons move the time indicator 8 steps at a time When the time indicator...

Page 18: ...d through the patient s tissue such as a finger or a toe to a receiver on the other side The sensor measurement wavelengths are nominally 660nm for the Red LED and 940nm for Infrared LED Maximum optic...

Page 19: ...t 2 Turn on the monitor The LCD screen will display the parameter monitoring screen 3 Attach the sensor to an appropriate site on the patient s finger 4 The readings will be displayed on the LCD scree...

Page 20: ...a xenon light source bilirubin lamps fluorescent lights infrared heating lamps or direct sunlight exposure to intense illumination can be corrected by covering the sensor with a dark material Excessiv...

Page 21: ...orrect results Prolonged and continuous monitoring may increase jeopardy of unexpected change of dermal condition such as abnormal sensitivity rubescence vesicle repressive putrescence and so on It is...

Page 22: ...Because the temperature sensor s impedance varies slowly with the temperature of the body an accurate result of temperature measurement is usually obtained after monitoring for more than 5 minutes 6 3...

Page 23: ...ing the functional safety check after the monitor has been used for 6 12 consecutive months or after monitor servicing or system upgrading This is to ensure the normal operation of the system Store th...

Page 24: ...h ppm depends on how much organic matter blood propagation grume etc existing on the surface CAUTION Never use strong solvent such as acetone ALWAYS dilute the solutions according to the manufacturer...

Page 25: ...r or any solution or pour water or any solution on the equipment ALWAYS wipe off all the excess liquids on the equipment surface and accessory surface with a dry cloth Never use EtO and formaldehyde t...

Page 26: ...ownload it from Internet and installed it in your PC Please refer to the download instruction and the HELP in SPDV for details NOTE The download and installation procedure maybe different depending on...

Page 27: ...ounds the alarm and the parameter s measurement flashes on the screen Prompting Icons of battery capacity pulse sound switch and alarm sound switch Power saving The monitor will automatic shut down wh...

Page 28: ...densing Altitude Operation 86 106 KPa Transportation and storage 50 106 KPa Parameter specifications SPO2 Patient Adult child Range 35 100 Resolution 1 Accuracy 70 99 2 0 69 Unspecified PR Range 30 25...

Page 29: ...lp but we cannot normally offer a refund DOA dead on arrival For additional clarification please refer to our terms and conditions at www daray co uk terms In a small number of cases we may determine...

Page 30: ...ray s instructions Attachment of fittings and accessories not approved by Daray Repairs modifications carried out by service personnel not approved by Daray Damage caused by chemical corrosion from cl...

Page 31: ...PURCHASED FROM PRODUCT SERIAL No ALTERNATIVELY REGISTER ONLINE AT WWW DARAY COM Freepost Plus RRAS YGXE SLBC Daray Ltd Marquis Drive SWADLINCOTE DE12 6EJ 1 YEAR WARRANTY Occasionally DARAY would like...

Page 32: ......

Page 33: ......

Page 34: ......

Page 35: ......

Page 36: ...TM VetZ Antimicrobial protection BioProtect Marquis Drive Moira Derbyshire DE12 6EJ Tel 0870 777 2664 Fax 0870 777 2665 email info daray co uk url www daray co uk...

Reviews: