DARAY SL720 Installation Manual Download Page 17

 

Page 16 of 17 

Service Notes 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Summary of Contents for SL720

Page 1: ...SL720 LED Mobile Minor Surgical Light Operating Installation Manual QAM SL720LFM 0219 4...

Page 2: ...Page 1 of 17...

Page 3: ...packing list 6 2 3 Pre Start Checks 6 3 Operation 7 3 1 Turning the light ON OFF and adjusting the intensity 7 3 2 Correct usage of the light 8 4 Maintenance 10 4 1 Maintenance schedule 10 4 2 Safety...

Page 4: ...patch to be directed wherever it is required for examination DARAY s BioProtect anti microbial coating is applied to all painted finishes including the light head and the handle All other surfaces ar...

Page 5: ...se visit www daray co uk or call 0333 321 0971 This product was designed manufactured in Great Britain by Daray Ltd Edison House Robian Way Swadlincote Derbyshire DE11 9DH 1 1 Supported Lights The Fol...

Page 6: ...djustable Colour Rendering Index CRI 85 93 Colour Rendering R9 56 90 Light Field Diameter D50 100 180mm Focusing depth 800mm Number of LED s 12 Power Consumption 30W Average Working Life 50 000 Hours...

Page 7: ...n appropriate and competent structural assessor The components supplied by DARAY for fitting installation should be the following items in the specified quantities SL720LFM Qty IEC Cable 5m Orange 1 M...

Page 8: ...uminated If the indicator does not illuminate check the power supply and then check the fuses located in the IEC socket on the mobile base see section 4 4 FUSE REPLACEMENT and also in the mains plug F...

Page 9: ...articular the flexible neck gooseneck If you need to bend the neck to reposition the light head keep one hand approximately one third up the neck and then use your other hand to curve the gooseneck th...

Page 10: ...the correct way to bend the neck ensure you curve it through its entire range This is the incorrect way to bend the neck Notice the sharp kink in the neck this will ultimately affect future flexibilit...

Page 11: ...lude Check all electrical connections are sound and that there is no visible cable wear All fixings to be checked and be of sound construction 4 2 Safety Precautions For all cleaning work power off th...

Page 12: ...products feature our BioProtect antimicrobial coating on painted finishes This is designed to assist with infection control and is not a replacement for cleaning CAUTION Our products are NOT waterproo...

Page 13: ...part of the IEC inlet mounted on the rear panel of the mobile base Unplug and remove IEC cable Using a flat bladed screwdriver or similar pull forward the drawer containing the fuses Hinge down the f...

Page 14: ...Power Cable 5m Hi Visibility D14658 All of the above parts can be purchase via DARAY For any SPARE PARTS or TECHNICAL ISSUES phone DARAY S Service Delivery Team on 44 0 800 878 9864 Opening times 9am...

Page 15: ...rges may be applicable OTHER Any situation which is not covered by any of the above We will always try to help but we cannot normally offer a refund For additional clarification please refer to our te...

Page 16: ...Daray s instructions Attachment of fittings and accessories not approved by Daray Repairs modifications carried out by service personnel not approved by Daray Damage caused by chemical corrosion from...

Page 17: ...Page 16 of 17 Service Notes...

Page 18: ...Page 17 of 17...

Page 19: ...Page 18 of 17...

Page 20: ...DARAY Ltd Edison House Robian Way Swadlincote Derbyshire United Kingdom DE11 9DH 44 0 800 878 9864 44 0 333 321 0973 sales daray co uk www daray co uk...

Reviews: