Page 21 of 35
4.2 Wall Mounted Version (EXCELW)
4.2.1 Considerations
Wall Construction
Before fixing it is important to consider the strength of
the wall and the depth of the fixings used.
Use of a building engineer or structural drawings may be
required to decide on suitability.
Examinations to be
undertaken
The Wall mounted version of the EXCEL has a
recommended install height where the bottom of the wall
bracket is approximately 1900mm from the floor.
It may be necessary to adjust this height, if there are
specific medical procedures the light will be used for.
If in doubt consult the end user or DARAY. When installing
the bracket take consideration of where you want the
lighthead to sit when in its lowest position.
Heavy or Difficult
fittings
Fitting the mounting plate can be difficult and aspects of the
fitting may require more than one person; this work should only
be attempted by appropriate and competent personnel.
Existing Environmental
Factors
When fitting, considerations have to be taken for existing
services such as; cabinets, vents, light switches, etc. along with
interior wall construction and any surrounding hazard.
Environmental
Responsibility
Once our products are installed all packaging should be disposed
of and recycled appropriately and in accordance with local
governing regulations.
Summary of Contents for EXCEL
Page 1: ...EXCEL LED Dental Examination Light Operating Installation Manual QAM EXCEL 0119 4...
Page 2: ...Page 1 of 35...
Page 7: ...Page 6 of 35 1 4 Technical Drawings...
Page 8: ...Page 7 of 35...
Page 13: ...Page 12 of 35 4 Installation Guide Section...
Page 36: ...Page 35 of 35 Service Notes...
Page 37: ...Page 36 of 35...
Page 38: ...Page 37 of 35...
Page 39: ...Page 38 of 35...