4.1.8 Fitting the lighthead
Connect the lighthead connector to the connector from the arm; apply a small
amount of grease to the light-head spigot and insert it into the arm, being careful
not to trap the cable/connector.
Whilst supporting the lighthead in one hand, insert and tighten the locking grub-
screw with the hex key provided.
Please Note: The grub screw should be as tight as possible.
Check everything for tightness and safety, then the arm and head for smooth and
correct freedom of movement; rebalance the suspension arm if necessary.
Once checked and fully earthed, the light can now be tested by applying power.
Press the power button on the back of the head or use the hand sensor located on
the front of the lighthead under the fascia.
Check the variable intensity by either holding your hand in front of the sensor or by
pressing and holding the rocker switch. Your new Daray light is now ready for use.
See section 3: Operation