Spyder3 S3-14 and S3-24 Monochrome Camera User's Manual
03-032-20117-01
Teledyne DALSA
Look-Up Tables
N ote:
This inform ation only ap p lies to the 4k m od el cam era.
The flat field corrections FPN and PRN U assu m e a linear resp onse to the am ou nt of light by the sensor,
ou tp u t nod e, analog am p lifier, and analog to d igital converter. To correct any non -linearity in this system
of com p onents a Look-Up Table (LUT) has been im p lem ented in the FPGA for each tap im m ed iately after
the ADC. The LUT ad d s a signed valu e (-256 to +255) ind exed by the 10 MSB of the inp u t valu e.
Calibrate Input LUT
Calibrates the cu rrent inp u t look -u p table for correcting non -linearity in the analog chain (CCD sense
nod e and analog-to-d igital conversion).
Camera Link Command
Parameter
Description
Notes
cil
0 to 1. 0 for d ark, 1 for light.
This com m and calibrates all
tap s w ithin the ROI.
To calibrate:
Place a w hite reference in front of the cam era. This is sim ilar
to a PRN U calibration.
In ad d ition:
Use the
w il
com m and to w rite the LUT to non -volatile
m em ory.
Use the
eil
com m and to enable u se of the LUT.
Use the
roi
com m and to lim it the tap s calibrated and to lim it
w hich p ixels are u sed for calibration. If a tap is not in the
region of interest, then it w ill not be calibrated and left at
cu rrent valu es.
Press sp acebar to abort this com m and .
Reru n this com m and if the analog gain or op erating
tem p eratu re changes.
Example
Enable Input LUT
Enables or d isables the u se of the inp u t look -u p tables for the correction of the analog ch ain non -linearity.
Camera Link Command
Parameter
Description
Notes
eil flag
0
Disable
1 Enable
Coefficients m u st be created first w ith the cil com m and .
Setting saved w ith the w fs and w u s com m and s.
Example