DALSA
Falcon XDR and HG Series Camera User's Manual
42
Serial Interface: How to Control the Camera
Mode 10: External Frame Rate, Fixed Exposure Time
In this mode, the frame rate is triggered externally using the EXSYNC falling edge. There is a 100
μ
s
delay between the trigger and the start of the exposure. The minimum exposure time is 10
μ
s. The
minimum delay between the readout and the next EXSYNC is 1
μ
s. Note, the exposure and readout time
cannot overlap.