ZX-3
Ultrasonic Thickness Gauge
27
The procedure to select the zero option only, is outlined below:
Zero (Auto/Manual)
1) Press the
key to access the menu items/features.
2) Use the
keys to scroll through the items/features until the
ZERO
feature is being displayed.
3) Press the
key to edit the zero setting. The edit icon will be illuminated
and flashing.
4) Use the
keys to select the auto/man option.
Note:
Pressing the
key prior to pressing the
key will abort to the
measurement screen without saving changes.
5) When the desired
ZERO
setting is displayed, press the
key to set the
status and return to the measurement screen.
6.6 Probe Diameter
The
PROB
feature was added to improve linearity when using a specific probe
diameter. The default standard setting is (.250”), and works reasonably well using a
general correction curve for all of our dual element transducers in our range.
However, selecting the exact diameter of the transducer being used will offer
additional linearity (accuracy). The three options found in our range of transducers
are (.187, .250, .500) inches. All of our transducer diameters and frequencies are
marked on top of the transducer housing.
The procedure to select the probe/transducer diameter is outlined below:
www.calcert.com
1.888.610.7664