PR-8V
High Performance Thickness Gauge
71
Note:
In the photos above, the MEM format is set to GRID. However, if the
ENGINE format was selected, the COL would change to CYL#, and the ROW
would be displayed as CLK#.
1) Press the
MEAS
key once to activate measure menu items. Press the
MEAS
key multiple times to move right and the
ESC
key multiple times to
move left until the
COL
or
ROW
cell is highlighted.
2) Press the
ENTER
key to display the
Grid View Box
.
3) Press the
UP
,
DOWN
,
LEFT
, and
RIGHT
arrow keys to scroll the stored
readings and corresponding display view. Notice as the cursor is moved to
a different cell, the display will be updated with the display view saved with
the reading.
4) The user may opt to clear a specific reading and save a new one at any
time. Press the
CLR
key in the appropriate cell location to clear the reading,
take a new measurement, and press the
OK
key to save the new reading.
5) Abort the
Grid View Box
by pressing the
MEAS
key at any time.
10.6 Deleting Grids (Files)
Deleting One or All Files from Memory
Deleting One File