PR-8V
High Performance Thickness Gauge
11
4) Press the
LEFT
and
RIGHT
arrow keys to scroll the digit locations.
5) Repeat steps
3 & 4
until the
B-ST
or
B-DEP
value is correctly displayed.
6) Press the
OK
key to set the
B-ST
and
B-DEP
value and return to the
measure screen, or
ESC
to cancel entering the
B-ST
or
B-DEP
value.
7)
Finally, press the
MEAS
key to return to the measurement screen and begin
taking readings.
Note:
The
B-ST & B-DEP
can also be adjusted from the menu tab items.
However, using the hot menu keys is the easiest method.
B-SCAN
DIGITS
In the upper left corner of each of the mode photos above, is the repeatability
indicator. The repeatability indicator is represented by six vertical bars and
represents how repeatable the measurements are. In regular measurement mode,
the
PR-8V
makes 4 measurements a second. In scan mode, the
PR-8V
makes 32
measurements a second. When the
PR-8V
is idle, only the left vertical bar and the
underline will be displayed. However, when the
PR-8V
is making a measurement,
five or six of the bars should be displayed on the repeatability indicator. If fewer than
five bars are showing, the
PR-8V
is having difficulty achieving a stable measurement
and the thickness value displa yed is potentially be unstable.