About DailyChek

®

 Premium SMBG System

Thank you for choosing DailyChek

®

 Premium Self-Monitoring Blood 

Glucose System. Designed for ease of use, the system can help you 
and your healthcare professional monitor and adjust your treatment 
plan to gain better control of your diabetes.

Important Information !

Expected Results for Non-Diabetic, Non-Pregnant Adults:
The normal fasting glucose range is 70 to 110 mg/dL (3.9 to 6.1 
mmol/L)
Two hours after meals, normal glucose values should be less than 
120 mg/dL (6.7 mmol/L)
Severe dehydration resulting from excessive water loss may cause 
false results.  If you feel you are suffering from severe dehydration, 
consult the healthcare professionals immediately.

• Test results below 3.9 mmol/L (70 mg/dL) mean low blood glucose 

(hypoglycemia).

• Test results greater than 13.3 mmol/L (240 mg/dL) mean high 

blood glucose (hyperglycemia).

• If you get results below 3.9 mmol/L (70 mg/dL) or above 13.3 

mmol/L (240 mg/dL), and do not have symptoms, repeat the test.

• If you have symptoms or continue to get results that fall below 3.9 

mmol/L (70 mg/dL) or above 13.3 mmol/L (240 mg/dL), follow the 
treatment advice of your healthcare professional.

Indicate the date you open the vial. Discard all unused strips 
immediately after 90 days from open date. The strips are for single 
use only.
If you are experiencing symptoms that are not consistent with your 
blood glucose test results and you have followed all instructions 
described in DailyChek

®

 Premium Self-Monitoring Blood Glucose 

System User’s Manual, call your healthcare professional.
Warning of potential biohazard: Healthcare professionals using this 
system on multiple patients should be aware that all products or 
objects that come in contact with human blood, even after cleaning, 
should be handled as if capable of transmitting viral disease.

Summary of Contents for HL568AA

Page 1: ...User s Manual Premium Self Monitoring Blood Glucose System Premium Blood Glucose Meter...

Page 2: ......

Page 3: ...g A warranty registration card is included with your system Please forward your warranty card immediately to HEALTH LIFE CO LTD If you have any concerns please call our HEALTH LIFE CO LTD office or au...

Page 4: ...ake them safe Do not block the openings of DailyChek Premium Blood Glucose Meters keep the openings free from dust hair stuff etc If you feel you are suffering from severe dehydration consult the heal...

Page 5: ...mber Use by in vitro diagnostic medical device AAA 1 5 V alkaline battery x 2 Do not reuse Temperature limitation Manufacturer Consult instructions for use Catalog number Serial number Biological Risk...

Page 6: ...em Key Features Get Ready for Testing Quality Control of the System Perform the Blood Glucose Test Use the Data Link Function Maintain Your Meter System Specifications Display Messages and Troubleshoo...

Page 7: ...mmol L 240 mg dL mean high blood glucose hyperglycemia If you get results below 3 9 mmol L 70 mg dL or above 13 3 mmol L 240 mg dL and do not have symptoms repeat the test If you have symptoms or con...

Page 8: ...e design DailyChek Premium SMBG Systems by using the latest biosensor technology It measures the glucose levels of the blood specimen by using a disposable dry reagent strip which can produce an elect...

Page 9: ...et Complete instructions 1 User s Manual 2 Quick Reference Guide 3 DailyChek Log Book 4 Warranty Card NOTE DailyChek Blood Glucose Test Strips and DailyChek Glucose Control Solution High and Low Level...

Page 10: ...ption DailyChek Premium Blood Glucose Meter Display Screen MODE Button Shows blood glucose results messages and glucose results stored in memory Press to enter the different setting mode which you wan...

Page 11: ...ymbol Indicates it s ready to apply the blood sample Decimal Point Appears in unit only set to mmol L Ketone Warning mmol L mg dL Temperature Symbol Indicates the ambient temperature is outside the op...

Page 12: ...irritation NOTE DailyChek Blood Glucose Test Strips are used with DailyChek Premium Blood Glucose Meter only Do not touch the strip with a wet hand Do not use strips beyond the expiration date printed...

Page 13: ...used with sterile lancet to obtain the blood for the testing Cap Depth Settings Indicator Lancet Ejector Release Button Cocking Handle Lancets It is used with lancing device to obtain the blood for th...

Page 14: ...ws users to add a food comment to the test result You may select the AC symbol if you perform the test before a meal or select the PC symbol if you perform the test after a meal It benefits you to man...

Page 15: ...etting for the Meal Comment Function is AC after meal testing where the AC symbol will appear with the test result If the AC comment is selected each test result will appear with the AC symbol Yet you...

Page 16: ...umulate in the blood and are excreted in the urine Ketone Warning When your test result is equal to or higher than 13 3mmol L 240 mg dL Ketone symbol appears on the screen It indicates there is a poss...

Page 17: ...old batteries To save the power please take out the batteries if you do not use the meter for a long period approx 2 months or more To avoid inaccurate results caused by electromagnetic interference...

Page 18: ...ng Mode If you power the meter on by loading batteries please skip this step The number of year is flashing Press SET button to scroll forward and select the correct number of year then press MODE but...

Page 19: ...ng Mode and follow the above step 6 to complete the code setting A Press SET button to select the correct number and MODE button to confirm the selection Then single digit starts flashing B Repeat the...

Page 20: ...strip correctly Fig 2 Inserting the strip with the wrong end will not turn on the meter Fig 3 Inserting the strip with up side down will not turn on the meter Fig 4 Inserting the strip with the wrong...

Page 21: ...en a new vial of test strips If you drop your meter If your result does not agree with the way you feel or result is lower or higher than expected Whenever you want to check for performance of the tes...

Page 22: ...ip vial you can go to next step If not refer to section Code Setting to set the code number Shake the glucose control solution well before squeezing Discard the first droplet and squeeze the second on...

Page 23: ...the expected range please check the following Are the control solutions or test strips within the expiration date Has the control solution been contaminated Is the code number correct Did you follow...

Page 24: ...nd be sure to put it back on to the lancet before you discard the lancet after the test Clean the outside of the lancing device by using alcohol or soap and water If the meter shows or symbol it means...

Page 25: ...6 Replace and screw the cap snugly Choose the depth of lancing Select 1 2 for soft o r thin skin 3 4 for average skin 5 6 for thick or calloused skin 4 Pull the cocking handle back until it clicks 5 T...

Page 26: ...de Setting to set the code number 8 Insert a new test strip make sure it goes all the way in The meter will turn on automatically when it detects the test strip 7 To avoid body fluid contamination wip...

Page 27: ...t on the top of the test strip 12 This strip has the correct amount of blood Make sure that the blood fills the reaction zone completely If too little blood enters the reaction zone an underfill an er...

Page 28: ...y from the strip After 5 seconds the device will beep again and display OK indicating the testing is completed 13 mmol L If the Meal Comment Function is activated on the test result will appear along...

Page 29: ...ip Or the meter will automatically shut down after 2 minutes if without any operation 17 Unscrew the cap of the lancing device and re cap the lancet then push the lancet ejector forward to remove the...

Page 30: ...1 LO and HI Readings Important information HI reading with symptoms If you feel symptoms such as fatigue thirst excess urination or blurry vision Then follow your doctor s recommendation to treat hyp...

Page 31: ...ults may not be as expected if the date setting of records is not precisely set The 7 day average is calculated from the test results obtained during the last 7 days counting from the date of last rec...

Page 32: ...tton the 14 28 60 90 day average record will appear in order You can then review the last 500 individual test result in memory 2 NOTE In the above figure the numeral 07d indicates the average value is...

Page 33: ...After the 90 day average the latest reading appears with the memory sequence number then the date and time instead Press MODE button to view the next most recent reading Then press MODE button each ti...

Page 34: ...delete all records Press SET and MODE buttons simultaneously for 3 seconds the screen will show dA for confirmation Press SET button again for 3 seconds to delete all records Or press MODE button to...

Page 35: ...Compatible computer with USB port CPU at least Pentium CD ROM drive CD ROM 4x or faster HDD 100MB or more space RAM 512MB or more Resolution of monitor 1024 x 768 Software requirements Compatible for...

Page 36: ...k Glucose Data Transmission Software User s Manual to transfer the test results in the meter memory to your computer Use the Data Link Function When USb appears on display it indicates that the meter...

Page 37: ...ther liquid into the test strip slot Never immerse the meter in water or any other liquid solution Display Check When the meter is off press any key to switch it on The display should look exactly lik...

Page 38: ...rature between 10 40 C 50 104 f Out of this range the system will not work Extreme humidity may affect the results A relative humidity greater than 80 may cause incorrect results Hematocrit The test r...

Page 39: ...rs Package with the lancet or strips Blood Volume 1 L Sample Type Capillary whole blood Hematocrit Hct 30 55 Power AAA 1 5V alkaline battery x 2 Battery Life At least 1000 measurements Display Dimensi...

Page 40: ...i p Follow instructions in Perform the Glucose Blood Test section Display messages What it means Troubleshooting M a k e s u r e t h e operation condition is within 10 40 C and R H 80 Repeat the test...

Page 41: ...e a or reading that m e a n s t h e t e s t e n v i r o n m e n t a l condition is out of desired range and the test will be terminated Repeat the test after replacing the battery See Get Ready for Te...

Page 42: ...tely DailyChek Glucose Control Solution Sold Separately HEALTH LIFE CO LTD 9F No 186 Jian Yi Road Chung Ho City 235 Taipei Taiwan Tel 886 2 8227 1300 Fax 886 2 8227 1301 E mail public hlmt com tw http...

Page 43: ...zhou China 215131 Tel 86 512 65799308 www sterilance com Meets the requirement of MDD 93 42 EEC Lancets Meets the regairement of MDD 93 42 EEC SteriLance Medical Suzhou Inc No 68 Litanghe Road Xiangch...

Page 44: ...ect in materials or workmanship HEALTH LIFE CO SERVICE will repair the meter or replace it with a new DailyChek Premium Blood Glucose Meter or equivalent product free of charge The warranty of the rep...

Page 45: ......

Page 46: ......

Page 47: ......

Page 48: ...P N 323101718 VER 001 20110518...

Reviews: