DAIHAN Scientific, Air-jacketed CO2 Incubators 8/14
■ How to switch from program/manual mode
Press and hold the PROGRAM / NORMAL button on the main
controller for 5 seconds to change the mode with the beep sound,
and the Program lamp will light. Press again for 5 seconds to
return to the previous mode
.
■ How to set program mode
In program mode (PROGRAM), each setting value can be
changed up to 10 STEPs. To change the set value of each STEP,
press the MODE button on the front panel and the first STEP
input screen appears. Temp (T), CO2, Door Temp (D) and Timer
(Tm) are displayed as shown on the left,
Using the number keys
and the SHIFT (←) buttons, enter the desired value and press
the ENTER button to the next set value. After all the settings are
completed, press the MODE button to go to the next step display
screen. At this point, setting the Timer value to 00:00 does not let
go to the next step but ends as previous step. Pressing the
MODE button displays the CYCLE input screen, which sets how
many times the programmed STEPs will be repeated.
Run Cycle No
:
This item sets how many times each
programmed STEP will be repeated. The maximum repeated
operation input value is 999 times, and when it is input as 000, it
repeats infinitely.
UV-LAMP Time
:
This item sets the time to turn on the UV LAMP
before starting the temperature control. After the UV lamp is
turned on for set time, temperature control starts.
Pgm-On Delay
:
It is displayed only when the control is not in
operation (STOP status). It is an item to delay starting of control
after set time.
Summary of Contents for ICA-100
Page 3: ...DAIHAN Scientific Air jacketed CO2 Incubators 3 14 1 3 Product Specifications...
Page 11: ...DAIHAN Scientific Air jacketed CO2 Incubators 11 14 1 7 How to replace HEPA Filter Option...
Page 17: ...DAIHAN Scientific Air jacketed CO2 Incubators 3 14 1 3...
Page 25: ...DAIHAN Scientific Air jacketed CO2 Incubators 11 14 1 7 HEPA Filter Option...