PIN # FUNCTION
REMOTE SET POS
1
Man Gain Pot
(+5V=Max; GND=Min)
Center
2
Man Black Pot
(+5V=Dark; GND=Light)
Center
3
Gain Switch
(+5V=AUTO; GND=MAN)
AUTO
4
AUTO EE
(+5V=ON; GND=OFF)
5
Gamma Switch
(+5V=ON; GND=OFF)
ON
6
Power IN
(+12V @ 200mA)
7
Black Switch
(+5V=AUTO; GND=MAN)
AUTO
8
Enhance Pot
(+5V=MAX; GND=MIN)
Center
Shield
GND
GATE
The gating pulse that the camera uses to
inhibit the readout of the CCD is brought in
through this connector. Gating cable 207663-
01 provides the interface. If a different
interface is required, a 4-pin male mini-DIN
connector, such as Dage P/N 737129-01 can be used. This
connector also allows the Dage
InvestiGater
to be used with
the CCD-100L to provide a complete system for long-term
gating. The camera mates to the
InvestiGater
through a
standard S-video cable such as Dage P/N 737980-01. The
pin-out and interface is as follows:
PIN # FUNCTION
1
GND
2
Power
In
(+12V @ 200mA)
3
Video
Out
4
External Gate In
(+5V=Readout; GND=GATE)
GATING
The CCD-100L can be externally gated in order to
provide additional sensitivity by inhibiting the readout of the