(2) Operation method for light emission 

 

(Picture screen 2) 

• When the sub power switch is 9ON:, normal figures are displayed as in the main screen of 

Picture screen 2. 

• If you touch the mode to use according to the purpose of the procedure, the curser is cre

-

ated, and if you want to change the settings, you can do so by touching the            

  part. 

 • Set and change the strength, pulse, time and memory of the laser output to use by touch

-

ing the       part. (time setting of 10 minutes recommended) 

 • If you touch the /START/part after fixing the Pad on the affected area by using the belt, 

light is released.  

(3) How to use the Pad on the part of the body. 

No

Procedure

Description

1

Check whether the Pad cable is properly connected 

to the equipment.

Multi Pad should be connected to 

Ch1~Ch4

2

Apply enough Accent PPC cream on the desired part.

2

Attach pad to the desired part such as abdomen and 

arms or legs by using band.

3

Two more laser probes are positioned onto the region 

of the lymphatic nodes closest to the treatment zone. 

[Refer. Lymphatic System]

4

Select desired frequency program by using Pulse 

mode of programs. [Refer. Protocol of LLL]

Frequency Range(Hz) : 

5, 50, 100, 500, 1000

5

Set operating time

Time(Min) : 5, 10 , 15, 20, 25, 30

6

Select desired output power by using POWER on the 

screen. [Refer. Protocol of LLL]

Output Power Range(mW) : 

80, 100, 120, 140, 160

7

Select the output CH that is attached on the body

CH1~CH6

8

Press the Start Button

* Cautions!   

* Warning! 

  

The patient must be held the safety switch if it is not comfortable during treatment.

V  

V   

V  

V   

As the light may affect the sight of the user when the device operates, the user never fails to 

wear the protective glasses, and the people around the user should wear protective glasses 

or other protective equipment. 

Appeared such as picture of left when you 

press the safety switch

Summary of Contents for LUMISLIM

Page 1: ...Doc No DYMLUMISLIM a Rev No 1 Date JUL 9 2010 LUMISLIM manual DAEYANG MEDICAL CO LTD...

Page 2: ...oved agents The Company will consider itself responsible for the effects on safety reliability and performance of the product only if assembly operations re adjustments modifications or repairs are ca...

Page 3: ...area or the area which is easy to getsplashed by water rain Please observe the recommended operation time and method do not use any other things than the product supplied by the manufacturer Install...

Page 4: ...ed Do not use a damaged electrical cord or aged receptacle There may be a risk of electrical shock or fire Do not give a shock to the product or accessories Do not put the electode to your eye directl...

Page 5: ...m setting simple and convenient How to save SAVE touch to or call CALL touch from the memory It is convenient if you use frequently used functions by saving them Random saving from SAVE1 SAVE7 availab...

Page 6: ...symptoms or who need to use a life support should confer with a doctor before using the product Use only certified safe accessory items Use only those accessory items listed in the equipment s instru...

Page 7: ...lasses to protect the eyes of the curer and patients 4 Fixing belt Belt fixing the Pad to the body 5 Power cord Electric wire connecting the power supply to the device User Interface No Name Function...

Page 8: ...ase in stall and use the specified order 2 Product set up 1 Unzip the storage bin inside the box Take out the unit PAD etc 2 Connect all of Pads to the respective jack on the front 3 Plug the power co...

Page 9: ...t PPC cream on the desired part 2 Attach pad to the desired part such as abdomen and arms or legs by using band 3 Two more laser probes are positioned onto the region of the lymphatic nodes closest to...

Page 10: ...he main screen Picture screen 2 the screen changes as in the Picture screen 5 If the desired CALL part is touched the screen returns to the main screen Picture screen 2 and the figures that have been...

Page 11: ...given from the outside of the device Keep the power cord separating it from the socket Keep the key separately in order to prevent any unauthor ized use by other people besides the manager Administra...

Page 12: ...Did you check the FUSE Plug in the input socket Check the Power switch Check the FUSE If possible persist unplug and call for ser vice GUARANTEE All DaeYang Medical units are guaranteed against defec...

Reviews: