10. Trouble shooting and correction

problem

cause

correction

Display

drifts

* Defective Sensor

* AC noise in power line

* Defective connection

to sensor (Noise)

* Input filter set incorrectly

* Check sensor state and output

*

Check

power

line

and

connection

*

Check

connection

to

sensor

* Check filter DIP switch

* 6-2) check display time

* 6-3) Change to AV mode

Display

19999 only

* Discordance between

sensor

output

and

DIP

S/W

setting

* Open and short of

sensor in/output line

* Check DIP switch about

sensor output

* Check sensor wiring

Display

change --

* Exchange of sensor

input line.

* data setting error

* Check sensor connection.

* refer to 6-3)

Not able to

select

FS,DP,CAL

* Data Protection

* Turn off data protection

refer to 3- 4)

Analog output

Problem

* Defective voltage and

current signal

* Difficulty of setting

* When 0

5V output is

required.

* Check DIP Switch 5

* Adjust zero and span VR

refer to 2-3)

* Connect 1/4W 500Ω

accuracy resistor between

GND and OUT port

* Contact the manufacturer.

Other noise

protection

* Connection noise filter with

AC input line

* Set noise filter to coil in parallel

* Check the shield.

* If you have any questions, please contact the manufacturer.

11. Other Models and Features

Features

Model

Signal

Excitation

Analog

output

Output

style

Data

hold

Peak

hold

User

Reset

SM-100

1 - 3mV/V

*0.5, 10 /V

DC 10V

*DC3V, DC5V

0 - 5,10V

0,4 - 20mA

Relay

*TR

O

O

O

SM-10H

1 - 3mV/V

*0.5, 10 /V

DC 10V

*DC3V, DC5V

0 - 5,10V

0,4 - 20mA

Relay

*TR

O

O

O

SM-15

1 - 3mV/V

*0.5, 10 /V

DC 10V

*DC3V, DC5V

-

-

-

O

O

SM-18

1 - 3mV/V

*0.5, 10 /V

DC 10V

*DC3V, DC5V

8 Step TR OPEN

Collector Output

-

-

O

SM-1B

1 - 3mV/V

*0.5, 10 /V

DC 10V

*DC3V. DC5V

B.C.D Output

TTL Level

O

-

O

SM-10R

0 - 5,10V

0,4 -20mA

DC10V,24V

*±15V

RS-232C Output

0-10V,4-20mA

Relay

*TR

O

O

O

SM-200

0 - 5,10V

0,4 -20mA

DC10V,24V

*±15V

0 - 5 , 10V

0,4 - 20mA

Relay

*TR

O

O

O

SM-23

4 - 20mA

*1 - 5V

DC24V

-

Relay

-

-

-

SM-25

0 - 5,10V

0,4- 20mA

DC10V,24V

* ±15V

-

-

-

O

O

SM-28

0 - 5,10V

0,4- 20mA

DC10V,24V

* ±15V

8 Step

TR OPEN

Collector Output

O

-

O

SM-2B

0 - 5,10V

0,4- 20mA

DC10V,24V

* ±15V

B.C.D Output

TTL Level

O

-

O

SM-20R

0 - 5,10V

0,4 -20mA

DC10V,24V

*±15V

RS-232C Output

0-10V,4-20mA

Relay

*TR

O

O

O

Common features : AC 110V or 220V 50-60Hz, Temperature Drift -- 0.02% FS/

,

A/D Sampling

15Times/Sec

Options : SAMPLING TIME -- 30Times/Sec

BCD Output MODEL with OPEN COLLECTOR CONVERT

* mark items are optional

* SM-10H : Sampling Time is 1000/sec

Head Office & Factory:

DACELL CO.,LTD

681-1 Cheoksan -Ri,Nami-Myeon,Cheongwon-Gun,

CHUNG-BUK, KOREA

TEL) +82-43-260-2242

FAX) +82-43-260-2245

Summary of Contents for SM110

Page 1: ... and ports short state current display value holding open state current value display normal operation mode 3 ANALOG OUTPUT GND Analog output is as following voltage or current DIP switch 5 on voltage output DIP switch 5 off current output ON Voltage Output 0 5V 1 5V 0 10V 2 10V OFF Current Output 0 10 0 20 4 20 Adjust zero point and maximum point of analog output using zero and span VR signal 0 z...

Page 2: ...ssible to adjust irrespectively Mode confirm and setting Unlock Press and key during 5 sec at the same time FS DP CAL LED will light up and reset lock mode data correction enable Lock Press and key during 5 sec at the same time in unlock mode FS DP CAL LED turn off lock mode power off Set lock mode automatically 5 AUTO ZERO mode Initialize sensor and indicator Execute this mode at unlock mode Exec...

Page 3: ...a is removed 1 x x x peak hold on only display and reform maximum measurement data peak hold off stored maximum measurement data is removed automatically mode display description Parity of Display x 0 x x standard type ex 20mA display 1000 0 x 1 x x display to signal and display indicate opposite ex 20mA display 760 0 Tracking mode of zero point x x 0 x standard type Display 0 trace minute signal ...

Page 4: ...ions please contact the manufacturer 11 Other Models and Features Features Model Signal Excitation Analog output Output style Data hold Peak hold User Reset SM 100 1 3mV V 0 5 10 V DC 10V DC3V DC5V 0 5 10V 0 4 20mA Relay TR O O O SM 10H 1 3mV V 0 5 10 V DC 10V DC3V DC5V 0 5 10V 0 4 20mA Relay TR O O O SM 15 1 3mV V 0 5 10 V DC 10V DC3V DC5V O O SM 18 1 3mV V 0 5 10 V DC 10V DC3V DC5V 8 Step TR OPE...

Reviews: