5. HARDWARE
1) Selection of AC input power
(1) Separate P.C.B from case.
(2) The following label is existed under power transformer
110V : connect 1 and 2
110
220
110
connect 3 and 4
220V : connect 2 and 3
2) Transformation of SAMPLING TIME
15 times/sec (standard).
(1) The following label is existed on the P.C.B
15 times/sec : connect 1 and C
○
○
○
15
C
30
30 times/sec : connect 2 and C
3) Setting of DIP switch
After connection, dip s/w must be set.
ON
1
2
3
4
5
6
7
8
1
mV/V
1.5
mV/V
2
mV/V
3
mV/V
voltage
1Hz 10Hz 100Hz
current
One of 1,2,3 or
4
turn
on
according
to
sensor
output
value (mV/V)
Turn
on
or off
according
to analog
output
analog input
filter
frequency
6,7,8
turn
off : 4㎑
6. SOFTWARE
1) CALIBRATION DATA
Set data as follows according to sensor output value.
Confirm calibration data after auto calibration execution.
1 mV/V range
---
0.700
1.249 mV/V
~
1.5 mV/V range
---
1.250
1.749 mV/V
~
2 mV/V range
---
1.750
2.499 mV/V
~
3 mV/V range
---
2.500
3.500 mV/V
~
2) Adjustment of display time and measurement data unit
(1) Press mode key, FS
DP LED turn on at the same time.
+
(2) 01.01 is displayed.
(3) Display time (set 1 times : fastest, set 15 times : slowest)
0
1
.
0
1
Setting of displat time
Setting of adjustment width
01 - After 1 times sampling, display
02 - After 2 times
“
:
:
:
:
14 - After 14 times
“
15 - After 15 times
“
01 : display 1, 2,
, 9
……
02
:
display
2,
4,
6,
8,
0(even)
05 : display 5, 10, 15
10 : display 10.....90
20 : display 20, 40, 60, …
50 : display 50, 00
3) PROGRAM
(SOFTWARE) OPTION
(1) Press mode key, DP
CAL LED turn on at the same time.
+
mode
display
description
Adjustment
of peak
hold mode
0 x x x
peak hold on : display maximum
measurement data
peak hold off : display present
measurement data..
reset : stored maximum measurement
data is removed.
1 x x x
peak hold on(only) : display and reform
maximum measurement data
peak hold off : stored maximum
measurement data is removed
automatically.
mode
display
description
Parity of
Display
x 0 x x
standard type.
ex)
20mA : display 1000.0
+
x 1 x x
display (
) to (
)
+
-
signal and display indicate opposite.
ex)
20mA : display -760.0
+
Tracking
mode of
zero point
x x 0 x
standard type.
Display 0 : trace minute signal to 0
x x 1 x
Display 0 : display minute signal.
Use in case of leakage measurement.
Average
value
display of
Signal
x x x 0
display measurement signal immediately.
x x x 1
display average value of measurement
signal (7 times sampling).
7. ALARM OUTPUT (RELAY OUTPUT)
High and low relay are a and b contact respectively.
Contact capacity is maximum AC 250V/5A and DC 30V/5A.
Relay contact operating time is maximum (1/Sampling Time) mm
Sec
* Output of transistor NPN open collector is option.
mode
setting
relay
HIGH
RELAY
high data > setting value
ON
high data
setting value
≤
OFF
LOW
RELAY
low data
setting value
≤
ON
low data > setting value
OFF
8. TECHNICAL DATA
item
data
allowed time temperature
40
80
―
℃ ~
℃
operating temperature
0
50
℃ ~
℃
power of sensor
5Vdc or 10Vdc
input signal type
1
3mV/V
~
max. load current
10Vdc/60
, 5Vdc/120
㎃
㎃
frequency response
Max 5㎑
analog output
0
10V,0
20
,1
5V, 4
20
~
~
㎃ ~
~
㎃
relay output
220V/5A (a,b contact)
relay response
Max 66
/sec
㎜
nonlinear characteristics
±0.02
/F.S
%
temperature coefficient
±0.02
F.S/
%
℃
display size
14
×8
㎜
㎜
sampling time
15 times/sec, 30 times/sec
filter frequency
1
, 10
, 100
, 4
㎐
㎐
㎐
㎑
short protection circuit
protection circuit (insert)
max. display range
-19999
19999
~
size
w 96
×h 48
×d 126
㎜
㎜
㎜
weight
650g
power
110Vac, 220Vac (50
60
)
~
㎐
9. Unit Label change
* Unit label is optional.
* In the case of exchanging label, refer to as follows
1, Separate P.C.B from the case.
2, Remove Conventional unit label and then adhere new
label.
3, Put together indicator in reverse process.