5. HARDWARE

1) Selection of AC input power

(1) Separate P.C.B from case.
(2) The following label is existed under power transformer

110V : connect 1 and 2

110

220

110

connect 3 and 4

220V : connect 2 and 3

2) Transformation of SAMPLING TIME

15 times/sec (standard).

(1) The following label is existed on the P.C.B

15 times/sec : connect 1 and C

15

C

30

30 times/sec : connect 2 and C

3) Setting of DIP switch

After connection, dip s/w must be set.

ON

1

2

3

4

5

6

7

8

1

mV/V

1.5

mV/V

2

mV/V

3

mV/V

voltage

1Hz 10Hz 100Hz

current

One of 1,2,3 or

4

turn

on

according

to

sensor

output

value (mV/V)

Turn

on

or off

according

to analog

output

analog input

filter

frequency

6,7,8

turn

off : 4㎑

6. SOFTWARE

1) CALIBRATION DATA

Set data as follows according to sensor output value.

Confirm calibration data after auto calibration execution.

1 mV/V range

---

0.700

1.249 mV/V

1.5 mV/V range

---

1.250

1.749 mV/V

2 mV/V range

---

1.750

2.499 mV/V

3 mV/V range

---

2.500

3.500 mV/V

2) Adjustment of display time and measurement data unit

(1) Press mode key, FS

DP LED turn on at the same time.

(2) 01.01 is displayed.

(3) Display time (set 1 times : fastest, set 15 times : slowest)

0

1

.

0

1

Setting of displat time

Setting of adjustment width

01 - After 1 times sampling, display

02 - After 2 times

:

:

:

:

14 - After 14 times

15 - After 15 times

01 : display 1, 2,

, 9

……

02

:

display

2,

4,

6,

8,

0(even)

05 : display 5, 10, 15

10 : display 10.....90

20 : display 20, 40, 60, …

50 : display 50, 00

3) PROGRAM

(SOFTWARE) OPTION

(1) Press mode key, DP

CAL LED turn on at the same time.

mode

display

description

Adjustment

of peak

hold mode

0 x x x

peak hold on : display maximum

measurement data

peak hold off : display present

measurement data..

reset : stored maximum measurement

data is removed.

1 x x x

peak hold on(only) : display and reform

maximum measurement data

peak hold off : stored maximum

measurement data is removed

automatically.

mode

display

description

Parity of

Display

x 0 x x

standard type.

ex)

20mA : display 1000.0

x 1 x x

display (

) to (

)

signal and display indicate opposite.

ex)

20mA : display -760.0

Tracking

mode of

zero point

x x 0 x

standard type.

Display 0 : trace minute signal to 0

x x 1 x

Display 0 : display minute signal.

Use in case of leakage measurement.

Average

value

display of

Signal

x x x 0

display measurement signal immediately.

x x x 1

display average value of measurement

signal (7 times sampling).

7. ALARM OUTPUT (RELAY OUTPUT)

High and low relay are a and b contact respectively.

Contact capacity is maximum AC 250V/5A and DC 30V/5A.

Relay contact operating time is maximum (1/Sampling Time) mm

Sec

* Output of transistor NPN open collector is option.

mode

setting

relay

HIGH

RELAY

high data > setting value

ON

high data

setting value

OFF

LOW

RELAY

low data

setting value

ON

low data > setting value

OFF

8. TECHNICAL DATA

item

data

allowed time temperature

40

80

℃ ~

operating temperature

0

50

℃ ~

power of sensor

5Vdc or 10Vdc

input signal type

1

3mV/V

max. load current

10Vdc/60

, 5Vdc/120

frequency response

Max 5㎑

analog output

0

10V,0

20

,1

5V, 4

20

㎃ ~

relay output

220V/5A (a,b contact)

relay response

Max 66

/sec

nonlinear characteristics

±0.02

/F.S

temperature coefficient

±0.02

F.S/

display size

14

×8

sampling time

15 times/sec, 30 times/sec

filter frequency

1

, 10

, 100

, 4

short protection circuit

protection circuit (insert)

max. display range

-19999

19999

size

w 96

×h 48

×d 126

weight

650g

power

110Vac, 220Vac (50

60

)

9. Unit Label change

* Unit label is optional.

* In the case of exchanging label, refer to as follows

1, Separate P.C.B from the case.

2, Remove Conventional unit label and then adhere new

label.

3, Put together indicator in reverse process.

Summary of Contents for SM110

Page 1: ... and ports short state current display value holding open state current value display normal operation mode 3 ANALOG OUTPUT GND Analog output is as following voltage or current DIP switch 5 on voltage output DIP switch 5 off current output ON Voltage Output 0 5V 1 5V 0 10V 2 10V OFF Current Output 0 10 0 20 4 20 Adjust zero point and maximum point of analog output using zero and span VR signal 0 z...

Page 2: ...ssible to adjust irrespectively Mode confirm and setting Unlock Press and key during 5 sec at the same time FS DP CAL LED will light up and reset lock mode data correction enable Lock Press and key during 5 sec at the same time in unlock mode FS DP CAL LED turn off lock mode power off Set lock mode automatically 5 AUTO ZERO mode Initialize sensor and indicator Execute this mode at unlock mode Exec...

Page 3: ...a is removed 1 x x x peak hold on only display and reform maximum measurement data peak hold off stored maximum measurement data is removed automatically mode display description Parity of Display x 0 x x standard type ex 20mA display 1000 0 x 1 x x display to signal and display indicate opposite ex 20mA display 760 0 Tracking mode of zero point x x 0 x standard type Display 0 trace minute signal ...

Page 4: ...ions please contact the manufacturer 11 Other Models and Features Features Model Signal Excitation Analog output Output style Data hold Peak hold User Reset SM 100 1 3mV V 0 5 10 V DC 10V DC3V DC5V 0 5 10V 0 4 20mA Relay TR O O O SM 10H 1 3mV V 0 5 10 V DC 10V DC3V DC5V 0 5 10V 0 4 20mA Relay TR O O O SM 15 1 3mV V 0 5 10 V DC 10V DC3V DC5V O O SM 18 1 3mV V 0 5 10 V DC 10V DC3V DC5V 8 Step TR OPE...

Reviews: