3. DISPLAY and KEY operation

1)DISPLAY :

4½ Digit LED

* Key operation start : press to

key.

Key operation exit : press to

key.

* Display : setting mode and operation mode.

1. Setting mode : data confirmation or correction.

2. Operation mode : normal operation state.

3. Press to

key : setting mode.

Press to

key : operation mode.

2) L.E.D Display

LED state

Description

HI

operation

Display operation mode of high relay.

setting

Confirm and adjust high data

LO

operation

Display operation mode of low relay.

setting

Confirm and adjust low data

FS

setting

Confirm and adjust full scale data.
Input the full range or capacity of sensor.

DP

setting

Confirm and adjust decimal point.

CAL

setting

key in sensor output data ( mV/V )
* refer to 6 - 1)

FS+

CAL

setting

auto calibration execution mode.
cal data is adjusted automatically.
* refer to 3 - 6)

F S +

DP

setting

Adjust display time and unit of

measurement data

* refer to 6 - 2)

DP+

CAL

setting

Software adjustment mode.
* refer to 6 - 3)

F S +

DP+

CAL

operation

Set or reset data lock mode
* refer to 3 - 4)

3) KEY operation

Mode
select

1. HI

LO

FS

DP

CAL

FS+CAL

FS+

DP

DP+CAL

2. All FND of selected mode will blink.

Positio
n
select

3. Press key : 10000→1000→100→10→1

Selected digit position will blink

4. DP(decimal point mode) : decimal point

position movement key.

10.000

100.00

1000.0

10000

Number
select

5. Data correction key.
6. 5 digit

1, -1, -, non

7. The others 1

2

3

4

5

6

7

8

9

→ → → → → → → →

0

Input

8. Input data. After setting up must press

key therefore become input.

A u t o
zero

9. Sensor : No load state(0V, 4

)

10. Display : 0

Key result

Description

Lock
mode

11. Protect setting data.

12. Reset lock mode to adjust data

4) PROTECTION mode

Use this mode for protection and security of System.
high, low data is possible to adjust irrespectively.

Mode

confirm and setting

Unlock

* Press

and

key during 5 sec at the same

time

FS, DP, CAL LED will light up and reset lock mode

(data correction enable)

Lock

* Press

and

key during 5 sec at the same

time in unlock mode : FS, DP, CAL LED turn off
(lock mode)

* power off : Set lock mode automatically.

5) AUTO ZERO mode

Initialize sensor and indicator.
Execute this mode at unlock mode.
Execution process (display : operation mode)

(1) Sensor : no load.
(2) Voluntary data display.

(3) Press

key and

key at the same time.

Zero display (ex : 0, 0.0, 0.00, 0.000)

(4) This data is memorized to indicator though power turn off.
(5) If operate mode with input signal to remain, this point is

memorized zero point.

6) AUTO CALIBRATION mode

Convenience to use when do not know calibration data of

sensor or displayed the measurement value differently.

EX)

At sensor Supply : 100.00 kg/㎠

Display

: 100.05 or 99.05 kg/㎠

Adjustment

: 100.00 kg/㎠

(1) Adjust zero point of indicator.
(2) Apply known load to sensor. Large load is high safety.

(3) Press

key : F.S and CAL L.E.D turn on at the same

time.

(4) As example table, 100.05 or 99.05 will blink.

(5) After correction of 100.00 using

key and

key,

press

key.

(6) 100.05 is display correction of 100.00.
(7) At this time, cal data is corrected minutely.
(8) If corrected cal data is over limits of 2-4), correct of DIP

S/W setting and again set the auto calibration.

4. Operating flow chart

Install and connection --Mount indicator to panel and make

connection. (Check AC POWER 110/220V)

Unlock mode -- Press

key(data protection unlock)

DATA Input

-- Set data(high, low, F.S, DP, CAL),

refer to 6

DIP S/W setting --sensor output data ---DIP S/W 1, 2, 3, 4

analog output

---- DIP S/W 5

filter frequency

---- DIP S/W 6, 7, 8

Auto zero

-- Sensor : no load

Press

key (auto zero)

AUTO CAL

-- Execute auto calibration.

Mode Check

-- Check function(Relay output,analog output,

reset, hold, peak).

Lock mode setting --Press

key(data protection mode)

Summary of Contents for SM110

Page 1: ... and ports short state current display value holding open state current value display normal operation mode 3 ANALOG OUTPUT GND Analog output is as following voltage or current DIP switch 5 on voltage output DIP switch 5 off current output ON Voltage Output 0 5V 1 5V 0 10V 2 10V OFF Current Output 0 10 0 20 4 20 Adjust zero point and maximum point of analog output using zero and span VR signal 0 z...

Page 2: ...ssible to adjust irrespectively Mode confirm and setting Unlock Press and key during 5 sec at the same time FS DP CAL LED will light up and reset lock mode data correction enable Lock Press and key during 5 sec at the same time in unlock mode FS DP CAL LED turn off lock mode power off Set lock mode automatically 5 AUTO ZERO mode Initialize sensor and indicator Execute this mode at unlock mode Exec...

Page 3: ...a is removed 1 x x x peak hold on only display and reform maximum measurement data peak hold off stored maximum measurement data is removed automatically mode display description Parity of Display x 0 x x standard type ex 20mA display 1000 0 x 1 x x display to signal and display indicate opposite ex 20mA display 760 0 Tracking mode of zero point x x 0 x standard type Display 0 trace minute signal ...

Page 4: ...ions please contact the manufacturer 11 Other Models and Features Features Model Signal Excitation Analog output Output style Data hold Peak hold User Reset SM 100 1 3mV V 0 5 10 V DC 10V DC3V DC5V 0 5 10V 0 4 20mA Relay TR O O O SM 10H 1 3mV V 0 5 10 V DC 10V DC3V DC5V 0 5 10V 0 4 20mA Relay TR O O O SM 15 1 3mV V 0 5 10 V DC 10V DC3V DC5V O O SM 18 1 3mV V 0 5 10 V DC 10V DC3V DC5V 8 Step TR OPE...

Reviews: