7. Calibration
Before calibration, You should select Dip switch to the most near thing in sensor
output value and make
.
7-1. Real-weight Calibration
Standard weight is necessary. When being used Multiple sensor at one Indicator, It
calibtrate certainly at Real-weight.
1) Real-weight Calibration Mode Setting
Press Function key about 3 seconds to return to Setup mode.
Press Up key to be
->
Press ENT key to choose calibration setting mode.
Preess Up key to be
Press ENT key to store it.
->
->
(In case it is not good,
will come out.)
If you want to get out of the setting mode and return to the measurement mode,
you may have to push the Function key.
(Currently the value is indicated.)
2) Zero and Span Setting
Press Up key about 3 seconds to return to Real-weight Calibration mode.
->
Leave the sensor free-load and press ENT key to set Zero.
->
->
->
Press Up key to be
Press ENT key, It uses
Shift key and
Up Key and it inputs Real-weight
value.
(Standard weight value it was prepared).
Press ENT key
, put counterweight on.
Press ENT key to set SPAN value.
14
->
->
->